Vitalograph micro - Instructions for Use
The flowhead may be checked with the daily calibration verification, to
be completed by the clinician/healthcare professional, and during the
periodic inspection of the device. Reference section 5.4. Calibration
Verification for details on how to check the device flowhead.
The clock will reset if the 3V coin cell battery has depleted. This will
be evident as the Time/Date option will be displayed every time the
device is turned on until the battery is replaced and the time/date is
set again.
The coin cell has an expected battery life of 15 years+, which exceeds
the devices product life; therefore a service is recommended at 10
years or when product useful life checks dictate it is required.
9. Customer Service
Service and repairs should be carried out only by the manufacturer, or
by Service Agents approved by Vitalograph.
For the names and addresses of approved Service Agents, please
refer to the contact information at the start of this manual.
Any serious incident that has occurred in relation to the device should
be reported to Vitalograph or its Authorised Representative and the
Regulatory Authorities of the country. Refer to the Vitalograph contact
information at the start of this document.
10. Consumables and Accessories
Cat number
Description
28501
Eco BVF (100)
28554
Eco BVF with Bite Lip (75)
28553
Eco BVF with Bite Lip + Disposable Nose Clip (75)
20303
Nose Clip (200)
28350
BVF (50)
36020
3-L Precision Syringe
69131
Flowhead Cone (5)
41653
USB Cable