Form A-755-6

 

Rev.: 4.0

 

Rev. Date:  July 10, 2016

 

 

 

 

 
 
 
 
 
 
 

 

Total-Lift Bed

 

 

 
 

Elevating  Patient Care  & Mobility

 

 

 
 
 
 
 
 
 

ASSEMBLY, OPERATING  AND  MAINTENANCE MANUAL

 

 
 

 

Bed  Model:

 

VG-TLB 425V/T

 

 

DEALER: 

This manual MUST be  given 

to  the  user  of  the  product.

 

USER: 

BEFORE using  this  product, read

 

this  manual and  save  for  future reference.

 

 
 
 
 
 
 
 
 
 

 

For  more information regarding 

VitalGo products, parts, and services, 

please visit 

www.vitalgosys.com

 

 

 
 
 
 
 
 

 

 

© 2015  Vitalgo Systems  Ltd. All rights  reserved.  Vitalgo, the  Vitalgo logo,  Total-Lift Bed™,  Joy Bath,  Joy Spa,  Total-Lift Chair  and  Total-Lift Toilet Seat  are  trademarks of  Vitalgo Systems  Ltd. Other 

names  are  the  property  of  their  respective  owners.

 

Содержание VG-TLB 425V/T

Страница 1: ...USER BEFORE using this product read this manual and save for future reference For more information regarding VitalGo products parts and services please visit www vitalgosys com 2015 Vitalgo Systems Lt...

Страница 2: ...eath or serious injury WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situat...

Страница 3: ...roperty damages Possibility of damage to motor material or environment OTHER SYMBOLS Useful tip For easier operation or better understanding of the unit Company name and address Manufacturing date IPX...

Страница 4: ...RVICE INFORMATION 10 RADIO FREQUENCY INTERFERENCE 11 WEIGHT LIMITATIONS 11 BEFORE PUTTING THE BED INTO OPERATION FOR THE FIRST TIME 11 WARNINGS FOR USERS AND CAREGIVERS 11 DEFINITIONS OF INVOLVED PERS...

Страница 5: ...RLENBURG POSITION IN BEDS WITH THIS FUNCTION 23 SIDE RAIL CONTROL PANEL OPERATION 24 CLEARING ERRORS 24 INITIALIZATION OF THE BED ONLY BY TECHNICIAN 25 STRAPPING SYSTEMS OPTIONAL 25 NON AMBULATORY PAT...

Страница 6: ...4 OPERATING INSTRUCTIONS 36 CPR FUNCTION VG TLB425 CAM4 MATTRESS SYSTEMS 38 TECHNICAL SPECIFICATIONS 38 PERFORMANCE SPECIFICATIONS 38 MECHANICAL SPECIFICATIONS 38 ENVIRONMENTAL SPECIFICATIONS 39 SAFET...

Страница 7: ...d this manual carefully before operating the bed This manual applies to normal use of the bed and assembly and maintenance It is also a practical reference guide Keep the instructions in a handy place...

Страница 8: ...otal Lift Bed can come with powered air mattress and high quality non powered mattress WEIGHT BEARING CONTROL The Total Lift Bed is the only real true Weight Bearing control device knowing the exact a...

Страница 9: ...cian to examine the unit and repair it Keep all moving parts including the main frame mattress deck head and foot sections and all drive shafts free of obstruction i e Blankets sheets heating blankets...

Страница 10: ...herwise the warranty is void VitalGo will not be responsible for any damage or injury that may result ELECTRICAL GUIDELINES DANGER Connect the bed directly to an electrical outlet Do not use extension...

Страница 11: ...s with all requirements of the guidelines for medical products It is classi fied as a Medical Equipment Electrical Operated Hospital Bed in accordance with standard IEC60601 1 2005 En 60601 1 2006 thi...

Страница 12: ...HNICIAN A person trained and authorized by Vitalgo to service the bed or is working for VitalGo s distributor and was trained by distributor with VitalGo s guidelines The technician is responsible for...

Страница 13: ...instructions for users and caregivers Let the technician instruct you in the safe use of the bed Before each use make sure that the bed is in proper working order Take care that no obstacles such as f...

Страница 14: ...Form A 755 6 Rev 4 0 Rev Date July 10 2016 VG TLB 425V T Bed Description 14 of 42...

Страница 15: ...d under the operating instructions laid out in this instruction manual Any other use is deemed to be inappropriate SPECIAL FEATURES HEIGHT ADJUSTMENT Error range see page 16 Electrical height adjustme...

Страница 16: ...70cm Smart BreakTM Casters are lifted from floor when not in Wheels position Each Side of bed has central locking break system one side Directional Lock On one caster for easy maneuvering of bed This...

Страница 17: ...Form A 755 6 Rev 4 0 Rev Date July 10 2016 17 of 42 40 95 104cm 37 8 20 67 52 5cm 87 87 3 95 10cm 7 90 20cm 96cm 85 82 218cm 1 25 3 18cm 223 5cm 87 40 222cm...

Страница 18: ...be adjusted by electric motors The horizontal height of the reclin ing surface can be adjusted and inclined All adjustment functions are controlled by a Wired hand set and basic controls are availabl...

Страница 19: ...surfaces are mostly powder coated and some are painted Some steel parts have black oxide coating and chrome nickel coating There is also use of aluminum The washable HPL are made of HPL High Pressure...

Страница 20: ...t activate Simply release the wired handset button permit a slight pause and then activate the next operation Before placing the bed into use test it by operating it through all phases of its operatio...

Страница 21: ...n and the 11 Backrest is raised to an Upright position it might be that the Leg rest will move down in order to avoid pressure on the back and pelvis E When reaching the end position of the function t...

Страница 22: ...button 8 is pressed the footboard will move until it s end position To stop movement stop pressing B Pressing buttons 7 8 together will move the footboard out In both cases when two beeps are heard th...

Страница 23: ...or the Leg rest and Foot Lifter in a different position 9 IMPORTANT NOTE If the Bed is in an angle of above 30 degrees the handset functions will not operate except for moving to standing or sleeping...

Страница 24: ...kage it will freeze In order to clear such error button 3 4 should be pressed together A short series of beeps will be heard Wait until beeping stops Try to operate bed If bed does not operate make fu...

Страница 25: ...to its highest position something is wrong Check all connections and redo steps 1 and 2 if the bed still does not work consult with VitalGo s authorised technician At the end of the initialization pr...

Страница 26: ...regiver in order to avoid any risk of rolling out of bed knee buckling and falling when in upright position The non ambulatory patient should be supervised by a caregiver when using the bed functions...

Страница 27: ...There are four straps at each side make sure that the Velcro is attached firmly to avoid movement of the mattress The Velcro straps 2 should be at tached on the PVC slates to prevent lateral movement...

Страница 28: ...f the controller Initialize Bed Technician Only Call Technician Do Not operate bed Bed movement does not stop Continues pressing on Hand Set Let go of switches Bed is producing unusual sounds burning...

Страница 29: ...welds for cracks Inspect the head and thigh and leg sections for bending warping or damage Check drive shaft and drive shaft connections for bending damage or excessive wear Inspect pull tubes and mo...

Страница 30: ...nfection In case the same person is using the bed a routine cleaning of the Bed is hygienically sufficient Disinfection of the Bed is necessary in case of contamination with infectious or potentially...

Страница 31: ...age to surface of coating of bed for which the company will not be responsible for When using disinfectants always Wear gloves Avoid skin contact Use closeable bottles with dosage pumps Use a clean to...

Страница 32: ...s not on press button 1 to start scale operation 3 Bring Bed to horizontal position Check that required bedding is on the bed but patient is not on the bed 4 Master Zero Press and hold button 3 for 4...

Страница 33: ...lay will show the amount of pressure on the foot board 2 Press again button Foot Board button 5 the display will show the percentage between the pressure on the Foot Board and the patient weight The a...

Страница 34: ...r to fasten the top sheet to the mattress base EXPLANATION OF SYMBOLS USED ON THIS DEVICE SYMBOL EXPLANATION POWER Turns unit On Off SOFT FIRM Up or Down key adjusts patient comfort pressures levels M...

Страница 35: ...igure Control unit has short circuit over voltage protection with single dual fuse S not shown in the picture Power Fail PF LED flashes to indicate power outage Low Pressure LP LED flashes to indicate...

Страница 36: ...ressure and the even num bered air cushions will deflate from desired patient comfort pressure to below 10 60 of set pressure in the first half of the DYNAMIC cycle and vice versa for the second half...

Страница 37: ...icroprocessor will activate an audiovisual signal to alert the care giver by flashing the amber LOW PRESSURE LED and turning on the buzzer Once the low pressure problem is fixed the audiovisual signal...

Страница 38: ...attress have lead free mercury free and latex Free components MECHANICAL SPECIFICATIONS SUPPORT SURFACE AIR CUSHIONS 70 denier urethane coated nylon R F welded liquid proof and washable Cal 117 pass B...

Страница 39: ...o nents such as manifold hose assembly air cushions top sheet and foam base are interchangeable and can be easily cleaned or detached for laundry SAFETY AGENCY APPROVALS The standard for safety of Med...

Страница 40: ...1 and the angle will be displayed The reading shots off automatically after a while to save battery NOTE the Inclinometer is calibrated and there is no need to use all other buttons which are for chec...

Страница 41: ...ANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE PRODUCTS MODIFIED...

Страница 42: ...ther it must be done in a way that ensures the safety of both the patient and the caregiver and with full respect for the patient s dignity Based on a unique combination of advanced technolo gies Vita...