background image

Instruction manual 

VitalFlex Elite LTC Bed 

Last update:03.16.2021 

Page 30 of 34 

 

 

 

 

 

13.6

 

Information on electromagnetic compatibility 

 

Guidelines and manufacturer's declaration - Electromagnetic emissions 

The care bed is intended for operation in an environment as specified below. The customer or 

user of the healthcare bed should ensure that it is operated in such an environment. 

Interference 

emission 

measurements 

 

Concordance 

 

Electromagnetic Environment - Guide 

HF emissions 

according to CISPR 

11 
(partly) 

 

Group 1 

The nursing bed uses HF energy exclusively for its 

internal function. Therefore, its RF emission is very 

low and it is unlikely that adjacent electronic 
equipment will be disturbed. 

 

HF emissions 

according to CISPR 

11 
(partly) 

 

 

Class B 

 

The nursing bed is suitable for use in all facilities 

and those directly connected to a public supply 

network that also supplies buildings used for 

residential purposes. 

 

Transmittance of 

harmonics according 

to IEC 61000-3-2 

 

Class A 

 

Transmissions of 

voltage fluctuations / 
flicker to 
IEC 61000-3-3 

 

Matches 

 

Содержание Elite LTC Bed

Страница 1: ...on manual Assembly and operation VitalFlex Elite LTC Bed Vital Mobility Medical Supplies Inc 130 Bass Pro Mills Dr Unit 62 Concord ON Canada L4K 5X2 info vitalmobility ca www vitalmobility ca Tel 905...

Страница 2: ...Last update 03 16 2021...

Страница 3: ...l of delivery and scope of delivery 13 6 Assembly and commissioning 14 6 1 Assembly of the erector optional 14 6 2 Commissioning 15 7 Functional description 16 7 1 Technical overview 16 7 2 Handset wi...

Страница 4: ...before each use If you change ownership please include this instruction manual 13 2 Technical data electrical 29 13 3 Technical data environment 29 13 4 Classification 29 13 5 Weights of the individu...

Страница 5: ...mfort All beds are carefully checked by our staff before delivery The healthcare bed delivered to you has left our building in perfect condition When you receive the healthcare bed the responsibility...

Страница 6: ...e user is convinced of the proper condition before using the healthcare bed including accessories 1 1 Explanation of the symbols used In these operating instructions important information is indicated...

Страница 7: ...he relevant member of the medical profession or the medical industry A person is also considered to be an operator if he keeps medical devices ready for use outside of health facilities in his company...

Страница 8: ...designed for the storage of adults with a body height from 150cm and a body weight from 40kg to max 200kg It is suitable for use in retirement homes nursing homes and in home care and may only be ope...

Страница 9: ...r experience for proper handling 4 Safety instructions 4 1 General safety instructions Possible potential dangers which may occur despite proper operation must be pointed out separately during the ins...

Страница 10: ...il element and the upper edge of the bed frame must be less than 12 cm When in use the side rails must not remain in a diagonal position Before moving the bed disconnect the mains plug from the socket...

Страница 11: ...nstruct each user on safe operation before initial use Inform the user of any hazards that may exist if the device is not handled properly Only instructed persons may operate the nursing bed This also...

Страница 12: ...egulations and followed these operating instructions To detect possible defects in time and to ensure safe use a technical check visual and functional check must be carried out by qualified personnel...

Страница 13: ...of 12 cm to 16 5 cm is suitable when using the undivided wooden side rails When using the split side rails a maximum mattress thickness of 17 5 cm may be used Make sure that the dimensions of the mat...

Страница 14: ...check that the delivery is complete You will receive a fully assembled care bed consisting of the following parts 1 Adjustable lying surface with mounted adjustment drives 2 Lifting frame with two mo...

Страница 15: ...ace Make sure that the locking cylinder pin engages in the recess of the erecting fixture Attention The erecting bracket must not be used outside the latching mechanism Engage the snap in cylinder pin...

Страница 16: ...e handle Production month Production year 6 2 Commissioning Connecting the nursing bed to the mains socket Insert the mains plug into the socket The mains plug should always be accessible so that in a...

Страница 17: ...7 Locking key for sister 8 Drive for electrical adjustment of the backrest and thigh support of the lying surface 9 Mechanical grid fitting for adjusting the lower leg support 10 Bed lifting frame wi...

Страница 18: ...ation Thigh adjustment electric infinitely variable 0 30 Lower leg adjustment mechanical via 5 point Rastomat 0 55 Electric height adjustment infinitely variable 380 800 mm Infinitely variable Anti Tr...

Страница 19: ...uards Do not press multiple keys at the same time as this may overload and damage the system 7 3 Locking function for handset There is a lock on the back of the hand control All electrical adjustment...

Страница 20: ...that it is securely engaged The side rails are only intended to prevent persons from falling out of bed Under no circumstances should you climb over them or lean on them 7 5 Operating the central brak...

Страница 21: ...e emergency lowering once the 9V block batteries must be replaced with new equivalent ones alkaline manganese battery type 6LR61 However the 9V block batteries should be replaced every 2 years even if...

Страница 22: ...s to be able to lower the backrest safely in an emergency Execution of mechanical emergency lowering The first user relieves the backrest before the emergency lowering by lifting the frame and holding...

Страница 23: ...the healthcare bed by hand with a damp cloth We recommend suitable cleaning and care products as cleaning agents for wooden and plastic furniture Household cleaners without ammonia and abrasives are a...

Страница 24: ...Disruption Possible cause Remedy No function Mains plug not plugged in Plug in the mains plug Lock function on handset activated Unlock the handset Handset not plugged in Insert the handset into the c...

Страница 25: ...responsibility and with verifiable compliance with the 2 error rate If during the electrical test it can be proven that an error rate of 2 the test cycle can be extended to a maximum of two years If y...

Страница 26: ...ils without break crack or other damage 2 9 Fixed position of the side rails in their fastening 2 10 Erector with grab handle and erector holder undamaged and no deterioration 2 11 Mains cable connect...

Страница 27: ...he handset locking device On Off OK 4 6 Check of the first error safety by means of an integrated locking box in the handset without complaint 4 7 Function of the side rails secure engagement 4 8 Side...

Страница 28: ...part of the annual inspection or after repair or before each re use of the bed Move the lock to switching position I or II First move all bed adjustments to a slightly raised position Blocking functi...

Страница 29: ...waste at the end of its service life For environmentally friendly disposal please contact your local authority or Vital Mobility 13 Technical specifications 13 1Technical data mechanical Safe working...

Страница 30: ...3 db A at a distance of 1m Reclining surface drive back knee 2xMD125 Limoss company Height adjustment drive 1xMD120 Limoss company 13 3Technical data environment Temperature range operation 10 C to 40...

Страница 31: ...ce Electromagnetic Environment Guide HF emissions according to CISPR 11 partly Group 1 The nursing bed uses HF energy exclusively for its internal function Therefore its RF emission is very low and it...

Страница 32: ...istance in meters m The a field strength of stationary radio transmitters should be less than the compliance level at all frequencies according to a field study Interference may occur in the vicinity...

Страница 33: ...ed power W Protective distance dependent on transmission frequency m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 10...

Страница 34: ......

Отзывы: