When to Replace:

Nozzle filters are made of a cellulose
material which prevents liquid from
entering the interior of the pipette. If
the filter becomes wet, the pipette will
not aspirate or dispense until the filter 
is replaced.

Catalog No. 9060-5009

For electronic macro volume model 100–5000µL

Package includes:

• 10 nozzle filters

• 1 plug

• Instructions

Electronic Macro Nozzle Filter
Replacement Instructions

N

NO

OT

TE

E:: 

Wear gloves for this procedure.

1. Place a macro pipette tip onto the nozzle

and push on ejector sleeve until click is
heard.  Grip tip firmly near the nozzle,
and turn it in a counterclockwise
direction to unscrew and remove nozzle
from pipette.

2. Slide off ejector sleeve and set aside. 

3. Remove the rubber plug at the tip end

of the nozzle by inserting a straightened
paperclip into the hole in the plug.  Use
the inserted end of the paperclip to pry
the plug out of the end of the nozzle.

4.  Insert paperclip into the threaded end

of the nozzle and push out the old filter.

Procedure:

Nozzle

Ejector Sleeve

MacroFilterReplacement  9/8/06  12:34 PM  Page 1

 

 

  

 

 