VistaLab Ovation 9057-2009 Скачать руководство пользователя страница 1

When to Replace:

Nozzle filters are made of a cellulose
material which prevents liquid from
entering the interior of the pipette. If
the filter becomes wet, the pipette will
not aspirate or dispense until the filter 
is replaced.

Catalog No. 9057-2009

For electronic single channel models 2–125µL 
& 5–250µL; adjustable models 10–100µL &
20–200µL; fixed volume models 21 to 200µL

Catalog No. 9057-3009

For electronic single channel model 25–1250µL;
adjustable model 100–1000µL; fixed volume 
models 201 to 1000µL

Package includes:

• 25 nozzle filters
• Tubing
• Filter insertion tool
• 3 rubber nozzle plugs

• Instructions

Nozzle Filter Replacement

for electronic single channel, adjustable 
and fixed volume models

Note:  Wear gloves

1. Remove the ejector sleeve.

2. Remove nozzle by placing the rubber

tubing provided onto the nozzle, and
unscrew it in a counterclockwise
direction.

3. Remove the rubber plug at the tip end

of the nozzle by inserting a
straightened paperclip into the small
hole in the plug.  Use the inserted end
of the paperclip to pry the plug out of
the end of the nozzle.

Procedure: