When to Replace:
Nozzle filters are made of a cellulose
material which prevents liquid from
entering the interior of the pipette. If
the filter becomes wet, the pipette will
not aspirate or dispense until the filter
is replaced.
Catalog No. 9057-2009
For electronic single channel models 2–125µL
& 5–250µL; adjustable models 10–100µL &
20–200µL; fixed volume models 21 to 200µL
Catalog No. 9057-3009
For electronic single channel model 25–1250µL;
adjustable model 100–1000µL; fixed volume
models 201 to 1000µL
Package includes:
• 25 nozzle filters
• Tubing
• Filter insertion tool
• 3 rubber nozzle plugs
• Instructions
Nozzle Filter Replacement
for electronic single channel, adjustable
and fixed volume models
Note: Wear gloves
1. Remove the ejector sleeve.
2. Remove nozzle by placing the rubber
tubing provided onto the nozzle, and
unscrew it in a counterclockwise
direction.
3. Remove the rubber plug at the tip end
of the nozzle by inserting a
straightened paperclip into the small
hole in the plug. Use the inserted end
of the paperclip to pry the plug out of
the end of the nozzle.
Procedure: