P a g e
|
16
Recharging Battery Instructions:
Only use VistaLab Lithium ion battery (Cat # 9020-4001). The battery must not be replaced by non-rechargeable
rechargeable batteries from other manufacturers.
WARNING:
Use of other manufacturer’s batteries or improper use of the instrument (e.g., short circuit, mechanical
damage, overheating, etc.) may cause the battery to explode.
Charge time = approx. 3 hours (from 0% to 100% battery)
Battery life = approx. 40 hours (dependent on frequency and intensity of use)
Conditions for recharging = 0
°
C to 40
°
C (32
°
F to 104
°
F), 20% to 80% relative humidity
Recharging Procedure:
1
Attach the proper outlet adapter to the power supply pack and insert into wall outlet.
2
Attach the wide end of the USB-C charging cord to the power supply pack.
3
Plug the USB-C end of the charging cord into the charging port of the ali-Q 2 (under the aliquot volume dial).
ali-Q 2 can be used when plugged into the charger.
Replacing the Battery/ Removing Battery (Hard reset)
:
1
Remove battery cover
: On the bottom of the ali-Q 2,
A) Remove the white sticker on the battery cover to expose battery cover latch hole.
B) Insert a flat, narrow tool into latch hole and press on latch while pushing cover forward.
2
Remove ba
tt
ery:
Grip the battery in one hand and the white connector in the other hand and gently pull
apart.
Do not pull by the leads!
3
Replace battery:
Plug (new) battery into white connector until it snaps-in. Insert battery into compartment.
Slide battery cover back into place.