Calibration Mode: 

In Calibration Mode, the user enters the % errors calculated in the initial Verification Procedure. Once these values are 

entered, the unit will calibrate automatically. 

NOTE: The upper LED indicates the current calibration status. Solid blue 

indicates factory calibration, while solid white indicates user calibration 

values have been stored.

 

 

Step 1 - Enter Calibration Mode 

1

 

Set the volume dial to BELOW 0.0 mL.  

2

 

Hold the ali-Q 2 upside down. 

3

 

Press and hold the aliquot button for 5 seconds. Three tones will sound to indicate entry into calibration mode and 

the lower LED will be 

quickly

 blinking 

blue

4

 

Release the button and return the unit to upright orientation. 

5

 

Watch for the lower LED to blink blue before proceeding. 

NOTE: Calibration mode can be exited, without saving changes, during step 2 or step 3 by quickly pressing and 

releasing the 

aliquot button

. This will be followed by 5 

red

 LED flashes to confirm exit.

 

Step 2 - Enter the % error at 0.5 mL 

After entering calibration mode, the lower LED will be 

quickly

 blinking 

blue

. This indicates that the unit is ready to 

accept the calculated error at 0.5 mL.  

1

 

Enter the 0.5 mL % error value, from the previously performed Verification Procedure by setting it on the volume 

dial. 

For example, if the mean error is 1.5%, set the volume dial to 1.5. 

 

2

 

If the ali-Q 2 is dispensing too high (+1.5% error), press the 

Aspirate

 button to 

save the changes. If the ali-Q 2 is dispensing too low (-1.5% error), press the 

Dispense

 button to save the changes. The 2 LEDs will alternate 

blue

 5 times, 

accompanied by a tone, to confirm entry.  

3

 

Wait until the 

lower LED

 is blinking 

blue

 before proceeding.

 

Step 3 - Enter the % error at 5 mL

 

The lower LED will now be 

slowly

 blinking 

blue

, indicating that the unit is ready to accept the calculated error at 5 

mL. Repeat the process outlined in Step 2 with the previously obtained 5 mL % error value. The 2 LEDs will alternate 

blue

 5 times, accompanied by a tone, to confirm entry

(Note: If the LEDs blink YELLOW, the calibration values are out of bounds (+/-10%) and the values will not be stored. 

Call our Technical Application Specialist at 914-244-4068 for assistance.)

 

If the LEDs blink GREEN, the unit has been successfully calibrated. The unit will now return to normal operation.

 

A post-calibration Verification should always be performed to confirm the changes were successful and the ali-Q 2 

accuracy should now be within spec. 

Reset to Factory Calibration:

 

The ali-Q 2 has been calibrated from the factory to ± %2 at 0.5 mL and 5 mL. The user can easily revert to the factory 

calibration at any time. To reset the unit: 

1

 

Enter calibration mode using Step 1 in the Calibration Mode section.  

2

 

Turn the ali-Q 2 upright. 

3

 

Press and hold the aliquoting button for about 5 seconds, until the LEDs begin to blink 

green

.  

4

 

Release the button. The LED will blink 

green

 5 times to confirm the factory reset has been applied. 

Содержание 2100-0003

Страница 1: ...ali Q 2 Series Operating Manual...

Страница 2: ...tem 9 Wall Mount 9 Kickstand 9 Technical Specifications 10 Calibration 11 In lab Calibration Protocol 12 Initial PerformanceVerificationWorksheet 12 Troubleshooting 14 Maintenance 15 Cleaning Decontam...

Страница 3: ...8 Use only VistaLab accessories and spare parts See Ordering Information on pg 18 Do not attempt to make any technical alterations Do not dismantle the instrument any further than is described in the...

Страница 4: ...arging port USB C N A 4 Aliquot button Nylon 5 Aspirate button Nylon 6 Dispense button Nylon 7 Battery Lithium ion Lithium Polymer 8 Battery compartment cover PC ABS 9 Nozzle Cover Nylon 10 Nozzle Ins...

Страница 5: ...ional part of a total amount of a solution or suspension verb To divide as a solution into equal parts Aspirate verb to draw by suction Dispense verb to expel liquid Serological Pipet pipet noun dispo...

Страница 6: ...Mode see next page the speed that the ali Q 2 aspirates and dispenses liquid is controlled by the touch sensitive aspirate and dispense buttons The amount of force used to push on the buttons determin...

Страница 7: ...the bottom gray dispense button until bottom of meniscus is at the desired volume graduation ATTENTION Do not fill the serological pipet over the max volume Severe damage can occur if any liquid ente...

Страница 8: ...as you will need Ex If you need to dispense 10 aliquots of 2 mL you will need to aspirate at least 20 mL of liquid 5 Aliquot Carefully position the tip of the full pipet over the first receiving vess...

Страница 9: ...has different sizes and geometries These geometric differences contribute to changes in the physical characteristics of the system ali Q 2 can sense the effect of these differences and will automatica...

Страница 10: ...racy Precision CV 3 0 mL 2 1 0 3 mL 3 3 With any other brand of serological pipet any size Dimensions 7 2 x 7 0 x 2 2 18 3 cm x 17 8 cm x 5 6 cm Weight 350g Volume Capacity Accepts 1 to 100 mL serolog...

Страница 11: ...ends sending ali Q 2 to VistaLab for calibration or performing verification at least once a year or whenever conditions and equipment change In lab Verification Adjustment Users can verify the accurac...

Страница 12: ...e initial current error at 0 5 and 5 0 mL and record the 2 numbers to use in the following calibration procedure Worksheet Determine the current accuracy of your ali Q 2 by taking the average of 4 gra...

Страница 13: ...dispensing too high 1 5 error press the Aspirate button to save the changes If the ali Q 2 is dispensing too low 1 5 error press the Dispense button to save the changes The 2 LEDs will alternate blue...

Страница 14: ...filter properly see pg 4 Replace nozzle insert Send ali Q 2 in for repair service Pipet does not stay in ali Q 2 Nozzle insert worn or damaged Replace nozzle insert NOTE Using Wobble not pipets will...

Страница 15: ...54 nm wavelength from UV sterilization lamps NOTE The effects of the UV exposure will likely cause discoloration over time Autoclaving NEVER AUTOCLAVE ALI Q 2 Only the nozzle cover nozzle insert and m...

Страница 16: ...the power supply pack and insert into wall outlet 2 Attach the wide end of the USB C charging cord to the power supply pack 3 Plug the USB C end of the charging cord into the charging port of the ali...

Страница 17: ...hnical Support to receive Return Authorization Unit will be inspected for defects and repaired replaced as necessary at the sole discretion of VistaLab Customer will be responsible for service charges...

Страница 18: ...terile 5 pack 9020 4006 Replacement Membrane Filter 0 45 m sterile 5 pack 9020 4008 Replacement Silicone Nozzle Insert 2 pack 9020 4010 Replacement Nozzle Cover 1 pack 9020 4011 Replacement Kickstand...

Страница 19: ...of their service life According to the Directive 2002 96 EC of the European parliament and of the Council on Waste Electrical Equipment WEEE of 27 January 2003 electronic equipment requires disposal a...

Страница 20: ...rical equipment for measurement control and laboratory use General Requirements CAN CSA C22 2 No 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use General...

Отзывы: