Calibration

 

ali-Q 2’s 

aliquot mode 

volume delivery is factory calibrated to aliquot the volume set on the aliquot volume dial to the 

factory specs, as listed on Technical Specifications section, in a state-of-the-art ISO 17025 accredited calibration lab 

under the following conditions: 

- Temperature: 20°C - 22°C (68°F - 71.6°F)

- Relative humidity: 45% - 75%

- Using 

Wobble-not 

serological pipet

- Using ISO 3696 Grade 3 purified water

NOTE: 

Using ali-Q 2 with different liquids more/less dense than water

 may 

slightly 

affect the 

accuracy 

of the 

dispensed aliquots. However, 

precision 

is always maintained with the ali-Q 2 thanks to its Intelligent Measuring 

System. To adjust the accuracy of your unit, see page 12. 

Calibration Verification

VistaLab recommends sending ali-Q 2 to VistaLab for calibration, or performing verification, at least once a year or 

whenever conditions and equipment change. 

In-lab Verification / Adjustment: 

Users can verify the accuracy of the dispensed aliquots in four ways: 

1

 

Gravimetric

 - Use a digital scale and water, determine the weight (in grams) of the  dispensed aliquot. See the 

Performance Verification document for gravimetric procedure at

 www.vistalab.com/support

 

2

 

Colorimetric

 - Use a spectrophotometer to measure the absorbance value of diluted dye in a solvent of controlled 

volume. Absorbance value is then converted into a volume (µL). 

3

 

Visual

 - Use an approved graduated cylinder (Class A, To Contain, calibration  certified). 

4

 

Visual

 (least reliable*) - Use the volume graduations on the serological pipet as a reference of how much liquid is 

being dispensed. 

*

Serological pipets in general have ±2% accuracy (at full scale) of their volume graduations, so this is the 

least accurate

way of verifying.

Содержание 2100-0003

Страница 1: ...ali Q 2 Series Operating Manual...

Страница 2: ...tem 9 Wall Mount 9 Kickstand 9 Technical Specifications 10 Calibration 11 In lab Calibration Protocol 12 Initial PerformanceVerificationWorksheet 12 Troubleshooting 14 Maintenance 15 Cleaning Decontam...

Страница 3: ...8 Use only VistaLab accessories and spare parts See Ordering Information on pg 18 Do not attempt to make any technical alterations Do not dismantle the instrument any further than is described in the...

Страница 4: ...arging port USB C N A 4 Aliquot button Nylon 5 Aspirate button Nylon 6 Dispense button Nylon 7 Battery Lithium ion Lithium Polymer 8 Battery compartment cover PC ABS 9 Nozzle Cover Nylon 10 Nozzle Ins...

Страница 5: ...ional part of a total amount of a solution or suspension verb To divide as a solution into equal parts Aspirate verb to draw by suction Dispense verb to expel liquid Serological Pipet pipet noun dispo...

Страница 6: ...Mode see next page the speed that the ali Q 2 aspirates and dispenses liquid is controlled by the touch sensitive aspirate and dispense buttons The amount of force used to push on the buttons determin...

Страница 7: ...the bottom gray dispense button until bottom of meniscus is at the desired volume graduation ATTENTION Do not fill the serological pipet over the max volume Severe damage can occur if any liquid ente...

Страница 8: ...as you will need Ex If you need to dispense 10 aliquots of 2 mL you will need to aspirate at least 20 mL of liquid 5 Aliquot Carefully position the tip of the full pipet over the first receiving vess...

Страница 9: ...has different sizes and geometries These geometric differences contribute to changes in the physical characteristics of the system ali Q 2 can sense the effect of these differences and will automatica...

Страница 10: ...racy Precision CV 3 0 mL 2 1 0 3 mL 3 3 With any other brand of serological pipet any size Dimensions 7 2 x 7 0 x 2 2 18 3 cm x 17 8 cm x 5 6 cm Weight 350g Volume Capacity Accepts 1 to 100 mL serolog...

Страница 11: ...ends sending ali Q 2 to VistaLab for calibration or performing verification at least once a year or whenever conditions and equipment change In lab Verification Adjustment Users can verify the accurac...

Страница 12: ...e initial current error at 0 5 and 5 0 mL and record the 2 numbers to use in the following calibration procedure Worksheet Determine the current accuracy of your ali Q 2 by taking the average of 4 gra...

Страница 13: ...dispensing too high 1 5 error press the Aspirate button to save the changes If the ali Q 2 is dispensing too low 1 5 error press the Dispense button to save the changes The 2 LEDs will alternate blue...

Страница 14: ...filter properly see pg 4 Replace nozzle insert Send ali Q 2 in for repair service Pipet does not stay in ali Q 2 Nozzle insert worn or damaged Replace nozzle insert NOTE Using Wobble not pipets will...

Страница 15: ...54 nm wavelength from UV sterilization lamps NOTE The effects of the UV exposure will likely cause discoloration over time Autoclaving NEVER AUTOCLAVE ALI Q 2 Only the nozzle cover nozzle insert and m...

Страница 16: ...the power supply pack and insert into wall outlet 2 Attach the wide end of the USB C charging cord to the power supply pack 3 Plug the USB C end of the charging cord into the charging port of the ali...

Страница 17: ...hnical Support to receive Return Authorization Unit will be inspected for defects and repaired replaced as necessary at the sole discretion of VistaLab Customer will be responsible for service charges...

Страница 18: ...terile 5 pack 9020 4006 Replacement Membrane Filter 0 45 m sterile 5 pack 9020 4008 Replacement Silicone Nozzle Insert 2 pack 9020 4010 Replacement Nozzle Cover 1 pack 9020 4011 Replacement Kickstand...

Страница 19: ...of their service life According to the Directive 2002 96 EC of the European parliament and of the Council on Waste Electrical Equipment WEEE of 27 January 2003 electronic equipment requires disposal a...

Страница 20: ...rical equipment for measurement control and laboratory use General Requirements CAN CSA C22 2 No 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use General...

Отзывы: