ThermoFlash
LX-26
User Manual – English V.15 200904
18
T
HE SCREEN DISPLAYS THE MESSAGE
LO
When using the ThermoFlash LX-26 the
message Lo can show on the screen.
The temperature analyzed is under the
measuring range selected, either less than 32°C
(90°F) in Body Mode or less than 0°C (32°F) in
Surface Temp Mode.
This message displays in various cases – please find below a list of
the main cases:
Reasons for LO
message display
Advice
Temperature reading hampered
by hair, perspiration…
Make sure that there is no
obstruction prior to taking a
temperature.
Temperature hampered by an
air flux.
Make sure there is no air flux as
this could interfere with the
infrared system.
Temperature readings too close
together, the ThermoFlash did
not have the chance to boot
itself.
Respect the pause of 15
seconds minimum between two
readings – 1 minute pause is
advised.
The measuring distance is too
far.
Please respect the measuring
distance (between 5 and 8 cm –
2 in and 3.14 in).