
BewellConnect - BW-OX1 - User Manual - 072017v2
12
- Do not sterilize the device in an autoclave, with ethylene oxide or by immersing it in liquid. The
device must not be sterilized.
7. TECHNICAL SPECIFICATIONS
Product name
BewellConnect®
My
Oxy
Model
BW-OX1
Classification
Classe II
Applied part type BF
IP22
Display type
LED screen
LED Specifications
Wavelength: Red = 660 ±3nm - Infrared = 905 ±10nm
Radiant power : Red = 3.2mW - Infrared = 2.4mW
Dimensions
1.3 x 1.3 x 2.4 in / 32 x 32 x 62 mm
Weight
1.41 oz / 40 g
Display and
measuring ranges
SpO
2
: Display range : 35 - 99% / Measuring range : 70 - 99%
Pulse: Display and measuring ranges: 30 - 250 bpm
Accuracy
SpO
2
: 70 - 100% : ± 3%
Pulse : ± 2 bpm
Resolution
SpO
2
: 1%
Pulse : 1 bpm
Normal working
conditions
Temperature : 41°F - 104°F / 5°C - 40°C
Humidity : 15-93%RH
Storage conditions
Temperature : -13°F - 158°F / -25°C - 70°C
Humidity : ≤93%RH
Atmospheric pressure : 86-106Kpa
Technology
Bluetooth 4.0 Smart Ready
Battery
Lithium 280mAh
Autonomy
400 readings
Automatic shut-
down
8 seconds
Bluetooth
transmission
distance
10m
Frequency
2400 - 2483.5MHz
USB
USB port 2.0
Accessories
provided
Pulse oximeter, instruction manual, protective pouch, USB cable
8. TROUBLESHOOTING
If one of the following issues arises when using your
My
Oxy, consult this trouble-shooting
guide. If the problem continues, please contact our customer service department at
www.support.bewell-connect.com.
ISSUE
POSSIBLE CAUSE
SOLUTION
Sp0
2
or PRbpm
do not appear
correctly
Your finger must be
correctly inserted.
Change the position of your finger and try again.
SpO
2
reading
too small to be
measured.
Try several times. If you have any doubts about
a problem related to the product, seek medical
attention to obtain an accurate diagnosis.
Sp0
2
or PRbpm
are unstable
Your finger may not
be inserted deeply
enough.
Change the position of your finger and try again.
The user’s finger is
trembling or the
user is moving.
Try not to move.
Impossible to
turn on the
device
Battery low
Recharge the device with the USB cable
Oximeter is damaged Contact customer service
Содержание Bewell MyOxy
Страница 1: ...BewellConnect BW OX1 User Manual 072017v2 1 BW OX1 MyOxy...
Страница 4: ...Year Month Day Serial number SN SN Serial number...
Страница 5: ...EN BW OX1 MyOxy...
Страница 15: ...BewellConnect BW OX1 User Manual 072017v2 15 SP BW OX1 MyOxy...
Страница 26: ......
Страница 27: ...BewellConnect BW OX1 User Manual 072017v2 27 FR BW OX1 MyOxy...
Страница 37: ...BewellConnect BW OX1 User Manual 072017v2 37...
Страница 42: ...BewellConnect BW OX1 User Manual 072017v2 42...
Страница 43: ...BewellConnect BW OX1 User Manual 072017v2 43 WA R R A N T Y...
Страница 50: ...BewellConnect BW OX1 User Manual 072017v2 50...