Quick Guide - suPARnostic® Quick Triage measurement with aLF Reader
aLF Reader set up
1.
Place the aLF Reader on a stable and level surface.
2.
Plug the power supply cable into the connector at the rear end of the aLF Reader.
3.
If needed: Connect the DYMO LabelWritter to the aLF Reader with the supplied USB cable.
Turn on the aLF Reader
4.
Turn on the aLF Reader by pressing the power button
Wait for the operating system to boot up.
The main menu will appear on the touch screen display.
Perform measurement and preparation of sample
5.
To start a new test, touch “New Test” on the touch screen display
.
6.
Two measuring methods are available, the suPARnosticQT and the suPARnosticQT20
i.
For suPARnosticQT leave the device for 20 minutes on the table for incubation before measuring in
the aLF Reader.
ii.
For suPARnosticQT20 insert the device in the aLF Reader, where it will incubate for 20 minutes
before measuring automatically.
7.
Using the internal 2D barcode scanner, scan the 2D barcode with the desired program,
suPARnosticQT
(manual reading) or
suPARnosticQT20
(automated reading), provided in the kit.
NOTE: keep the barcode in a vertical angel 10
–
15 cm from the aLF Reader.
8.
Test Name, Lot ID, and Sample Settings will automatically appear on the screen.