Visitor T30 C-DR
User's Manual
[File: 218899-01-01-B.docx]
Page 29/74
5
MESSAGES
The equipment foresees three types of messages on display:
S = Status of equipment
No influence on the working
W = Warning
Press the key RESET on the keyboard to delete the warning message and go on to work.
F = Error
The equipment must be switched ON again.
A list of errors is created to ease the work of Service personnel.
The message is displayed in the language used for the
equipment configuration.
All warning messages must be reset by the operator
through the key RESET near the message.
Status messages
Typ.
Text
Meaning
Intervention
S
READY
The equipment is ready to perform an
exposure
S
WAITING
Preparation phase
Wait for message “READY”
S
MANUAL
S
DAP READY
Dosimeter ready
-
Warning messages
Text
Text
Meaning
Intervention
W
CLOCK ERR.
System clock error
Press the RESET key to go on
W
INI. APR
APR checksum error
Press the RESET key to go on
W
ERROR IN APR
An APR error is out of scale
Set differently the parameters
W
TUBE SEASONING
After a long idle period (3 months or
more) it is necessary to proceed to the
X-ray tube seasoning, with the
purpose to avoid serious faults
Press the RESET key to go on, call
Service for the tube seasoning
W
HOT TUBE
The temperature of the monobloc has
reached the max. allowed value
Wait the monobloc cooling
W
EXPIRED TIME
The x-ray handswitch has been
pressed at the “1° step” for more than
15 seconds
Release the handswitch and repeat the
x-ray
W
MANUAL STOP
The x-ray handswitch has been
released before the end of the
exposure
Press the RESET key to go on
W
INACTIVE DAP
Dosimeter not connected
-
-
MAX DOSE
The doses counter has reached the
max. value that can be displayed.
Press the F1+RESET key to reset the
value.
W
DAP ERROR
Dosimeter connected, but in error.
Press the RESET key and call
Service.
-
DAP RESET
The sum of the product-area doses
has been reset.
-