background image

READ AND SAVE THESE INSTRUCTIONS

Thank you for purchasing the Vicks SmartTemp Digital Thermometer, a professionally accurate instrument for fast and 

easy temperature taking. Using Bluetooth

®

 technology, the thermometer wirelessly connects to your Smartphone device to 

display temperature readings. Please read these instructions carefully to ensure accurate temperatures and safe operation. 

When used properly, this high quality product will give you years of accurate, dependable service.
This thermometer is intended for the intermittent measurement and monitoring of human body temperature orally, rectally, 

or under-the-arm. It can be used on people of all ages except pre-term babies or very small (for gestational age) babies. It 

is intended for household use only. This thermometer is intended to be used with Apple and Android mobile devices. Please 

see www.vicksthermometers.com/smarttemp for the full list of supported devices. If you have any questions or comments 

about the Vicks SmartTemp Digital Thermometer, call our toll-free Consumer Relations line: 1-800-477-0457  

or email [email protected].

WARNINGS AND PRECAUTIONS

•  This product is not intended to diagnose any disease. Use of this thermometer is not intended as a substitute for 

consultation with your physician

•  High, prolonged fever requires medical attention. Please consult your physician immediately.
•  Do not allow children to take their temperatures unattended. Keep the thermometer and all parts out of reach of children 

under 12 years.

•  Do not allow children to run or walk during temperature taking.
•  Never immerse the thermometer into water or other liquids.
•  Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the 

user’s authority to operate the equipment.

BEFORE USE

Your Vicks SmartTemp Digital Thermometer is a 3-in-1 thermometer for oral, rectal or underarm use. Whichever method 

is used, it is recommended that for 15 minutes prior to use, avoid eating or drinking any liquids, exercising, smoking, or 

taking showers or baths.

OPERATING YOUR  Vicks SmartTemp DIGITAL THERMOMETER

SETTING UP YOUR THERMOMETER AND SMARTPHONE FOR TEMPERATURE TAKING 

Note: This setup is only required once per Smartphone device. You do not need to perform this setup every time you take 

a temperature. 

1.  Download the Vicks SmartTemp App from the iTunes App Store or Google Play Store. The App can also be downloaded 

by going to www.vicksthermometers.com/smarttemp.  

2.  To pair the thermometer to your Smartphone, enable the Bluetooth

®

 setting on your phone and open the Vicks 

SmartTemp App.

3.  Following the prompts on your phone, PRESS AND HOLD the On/Off button on the thermometer until you see the 

STEADY RED LIGHT (approximately 5 seconds). When the STEADY RED LIGHT appears, press “Confirm” on the App to 

complete the setup.

Note: If the setup prompt does not automatically appear, go to the “More” (menu in the SmartTemp App and select “Add 

new device”.  Follow the prompts.

SETTING UP A PROFILE (

OPTIONAL

)

1.  Open the Vicks SmartTemp App on your mobile device.
2.  Select “Create Profile” and fill in the patient’s name, birthdate and photo (optional). Select “Save” when complete.
TAKING A TEMPERATURE

1.  Open the Vicks SmartTemp App on your mobile device and make sure the thermometer is On (Press On/Off button). 

If you have completed the initial setup, and the Bluetooth

®

 setting is enabled on your phone, the Smartphone and 

thermometer will connect automatically within 5 seconds.

3.  On the App, select the patient’s profile or select “Take a temperature without profile”. 
4.  Select the location where you intend to take a temperature (oral, rectal, underarm).
5.  Insert the thermometer. The display on your Smartphone will change to show the temperature rising, indicating that the 

thermometer is working.

6.  When the measurement completes, a beep will sound on the thermometer and your mobile device screen will display 

the temperature value. On this screen, you can click “+” to add details such as symptoms, medications, and notes.

7.  To retake, press “Retake temp” on your mobile device.

CLEANING INSTRUCTIONS

•  Probe covers can be purchased and used on this thermometer to prevent the spread of germs.
•  Clean the probe tip after each use by washing the tip with soap and warm water or by disinfecting with rubbing alcohol 

(70% isopropyl).

•  Wipe dry with a soft cloth.
•  Do not submerge the thermometer in water. It is not waterproof. DO NOT BOIL OR CLEAN IN DISHWASHER. This 

will cause the thermometer to no longer function and will void the warranty. Do not wash the unit with any thinner or 

chemical solvent.

CHANGING THE BATTERIES

The Vicks SmartTemp thermometer includes a long lasting battery. When the battery needs replacing, a message will 

appear on your mobile  

device indicating that the battery needs replacing.
1.  Using a coin, unlock the battery cover and carefully remove the battery with a non-metallic instrument
2.  Replace the battery with a CR2032. Insert the new battery with the positive side (+) facing the battery cover.  

Replace cover and close.

ERRORS AND TROUBLESHOOTING

For errors & troubleshooting, see the MORE menu on the Vicks SmartTemp App.

Vicks

®

 SmartTemp

 

Thermometer

Model VDT985US

On/Off 

Button

Flexible Probe

LED Light

Storage Case

Probe Tip

SPECIFICATIONS

Battery Life: 

More than 100 measurements or approximately 2 years if used every week.

Temperature Range: 

89.6°F-109.2°F (32.0°C-42.9°C)

Operating Ambient Range: 

50-104°F (10-40°C)

Storage Temperature Range: 

-13°F to 131°F(-25°C to +55°C)

Operating and Storage relative humidity: 

15-95% RH±5%

Accuracy for displayed temperature range:  ±0.2°F (±0.1°C) within measurement range 95.9-107.6°F (35.5-42°C),  

and ±0.4°F (±0.2°C) outside of this range

Compatibility:  

For use with Apple and Andriod devices. See www.vicksthermometers.com/

smarttemp for the full list of supported devices.

TEST PROTOCOL (Laboratory Use Only)

1.  Once thermometer is paired with Smartphone, launch the Vicks app.

2.  Go to the “More” tab and tap “Settings.” Then, tap “Measurement Mode” and select “Monitor Mode.”

3.  Go to the “Take Temp” tab and follow the prompts to check for device accuracy.

This thermometer is specified to operate at 1 atmospheric pressure (700-1060hPA) or at altitudes with an atmospheric 

pressure up to 1 atmospheric pressure (700-1060hPA). This appliance conforms to ASTM E 1112 (Kaz USA, Inc., 

Southborough, MA). This device complies with part 15 of the FCC Rules. Operation is subject to the following two 

conditions: 1. This device may not cause harmful interference, and 2. This device must accept any interference received, 

including interference that may cause undesired operation. For a complete list of standards visit the SmartTemp App or 

www.vicksthermometers.com.  Internally powered ME equipment.
This device is calibrated at the time or manufacturing.  No recalibration is required.  Contact Customer Service for any 
accuracy issues.

PERFORMANCE

•  ASTM E1112
•  EN 12470-3
•  ISO 80601-2-56

SAFETY

•  ISO 14971:2012
•  IEC 60601-1:2005

EMC

•  FCC Part 15 for Intentional Radiators
•  Category 1 Radio Equipment RSS-210 Bluetooth SIG Qualification
•  IEC 60601-1-2-2007 Electromagnetic compatibility series of standards

SOFTWARE

•  EC 62304:2006
•  FDA Guidance document: Radio Frequency Wireless Technology in Medical Devices

DESIGN

•  MEDDEV 2.1/6 (2012) – Guidelines for the qualification and classification of standalone SW used within healthcare  
  within the regulatory framework of medical devices
•  IEC/TR 80002-1:2009 – Medical device SW - Part 1: Guidance on the application of ISO 14971 to medical device SW
•  FDA Mobile Medical Applications
•  NB-MED/2.2/Rec4 – SW and Medical Devices (2001)
•  FCC part 18
•  Category 1 Radio Equipment RSS-210 and Industry CA Requirements for Intentional Radiators

ONE YEAR WARRANTY

Your Vicks SmartTemp Thermometer is durable and manufactured for normal, household use. Following directions carefully 

will ensure years of dependable operation. If for any reason (other than misuse or normal battery replacement) you are 

dissatisfied with your Vicks SmartTemp  Thermometer, we will repair or replace it, at our option, free during the one year 

warranty period. Modification to the product by the consumer is not authorized and voids the warranty.
If the thermometer does not function properly, first check the battery. Replace it if necessary. If repair is necessary call 

our Consumer Relations Department at 1-800-477-0457 or write to Kaz USA, Inc. Consumer Relations Department, 

250 Turnpike Road, Southborough, MA 01772 USA, or e-mail [email protected], explaining just how the 

thermometer is improperly working. Our consumer service representative will provide further instruction on how to correct 

the trouble yourself or will ask you to return the thermometer for repair or replacement.

k

 Manufacturing Quality Healthcare Products for Over 75 Years

 Kaz USA, Inc. • 250 Turnpike Road • Southborough, MA 01772 • www.vicksthermometers.com

This product is distributed by Kaz USA, Inc. under license from The Procter & Gamble Company.  

Vicks and other associated trademarks are owned by The Procter & Gamble Company. 

The Bluetooth

®

 word mark and logos are registered trademarks owned by Bluetooth SIG, Inc., and any use of such marks by Kaz 

USA, Inc. is under license.

16JUL15 

IP34

© 2015 Kaz USA, Inc.

To protect the environment, dispose of this product 

and empty batteries at appropriate collection sites 

according to national or local regulations.

ID #2ABRGVDT985

Made and 

printed in 

China

Отзывы: