Startup Max System Gain
and the
Running Max System Gain
to 2 times this value.
The
Startup Output Threshold
parameter
is the base output voltage when a test is starting.
The
Running Max System Gain
is a safety limit that is used to shut down the system when something
goes wrong (e.g. an accelerometer cable falls off). This value typically is the same as the
Startup Max
System Gain
parameter but may need to be increased if your shaker requires widely varying drive
levels at different frequencies. If a test is incorrectly aborted due to a Max Run System Gain error,
increase this value. The
Running Max System Gain
parameter can be determined by running a
sweep through the desired frequency range. After completing one sweep, find the highest output
voltage on a System Gain vs. Frequency graph. Enter 2 times the maximum value for the
Running
Max System Gain
parameter.
The
Running Max System Output
parameter is a safety limit that is also used to shut down the
system when something goes wrong. This parameter typically is double the
Startup Max System
Output
but may need to be increased if your shaker requires widely varying drive levels at different
frequencies. If a test is incorrectly aborted due to a Max Run Drive error, increase this value. The
maximum output voltage required for a test can be determined by running a single sweep of the test.
After completing one sweep, find the highest output voltage on the Output Drive graph (the cursor
display function may be helpful in finding this value.) Enter 1.5 times the maximum value for the
Running Max System Output
parameter.
2.5.2
How to tune Random controller parameters
How to tune Random controller parameters
Click the "Settings" button
on the toolbar, or select the
menu
item.
tab.
Start with the following values:
Starting Average
3
Starting Gain
0.1
Startup Time
15 sec
Running Average
20
Running Gain
0.05
Slew rate
20 dB/sec
Sample rate
Auto
Lines
800
Shutdown rate 60 dB/sec
Clipping6x Sigma
Window type
Blackman
There are two sets of control parameters:
·
The Starting parameters are used while the output is ramping up to the control level and generally
have less averaging and a higher gain value because in this mode the system must respond
quickly to equalize the response. A smaller
Starting Gain
will increase the time required to reach
the desired output level.
·
The Running parameters are used once the output has reached the desired level and generally
have higher averaging and lower gain values than the corresponding Startup parameters because
110
154
Содержание VibrationVIEW
Страница 15: ...Document Revision Part I ...
Страница 17: ...How To Part II ...
Страница 96: ...Menu commands Part III ...
Страница 142: ...Test Definition Part IV ...
Страница 217: ...Control Centers Part V ...
Страница 226: ...VibrationVIEW Help 213 1995 2005 Vibration Research Corporation All rights reserved See Also RecorderVIEW 309 ...
Страница 227: ...Accelerometer Toolbars Part VI ...
Страница 231: ...Stop Codes Part VII ...
Страница 263: ...Typical Views Part VIII ...
Страница 265: ...Typical Views 252 1995 2005 Vibration Research Corporation All rights reserved 8 4 Random Random CRT View ...
Страница 268: ...VibrationVIEW Help 255 1995 2005 Vibration Research Corporation All rights reserved ...
Страница 269: ...Dialog Boxes Part IX ...
Страница 303: ...Windows Elements Part X ...
Страница 309: ...APPENDIX Part XI ...
Страница 311: ...APPENDIX 298 1995 2005 Vibration Research Corporation All rights reserved ...