Страница 1: ...lance SJ CE Series Operation Manual IMPORTANT To ensure safe and proper use of the balance please read this manual carefully After reading this manual store it in a safe place near the balance so you...
Страница 2: ...ser friendly keys Furthermore the large liquid crystal display provides excellent visibility and the instrument s high speed and stability intrinsic to a tuning fork design help boost operational effi...
Страница 3: ...urement 15 6 Counting Pieces 17 6 1 Sampling 17 6 2 Increasing the Counting Accuracy Memory Update Method 19 7 Measuring Percentage 20 8 Limit Function 22 8 1 Limit Function Setup 22 8 2 Setup of Limi...
Страница 4: ...This term indicates steps that the user should take to ensure the quality and reliability of the balance Meanings of Symbols Each symbol is accompanied by an instruction Indicates a mandatory action t...
Страница 5: ...ation were it may be subjected to excess moisture Electric shock or short circuiting could occur The balance may be corroded with resultant malfunction 0 Do Not Leave Afloat Do not use the balance wit...
Страница 6: ...ate measurement Unplug Adapter If the balance is to be unused for an extended period of time unplug the adapter This conserves power and prevents deterioration 0 Do Not Use Do not use volatile solvent...
Страница 7: ...ee page 6 Level For checking the level of the balance Turn the adjusters until the bubble rests in the center of the red circle Operation keys See page 7 Front View round pan type RS232C connector Opt...
Страница 8: ...Stays onandflashes while span adjustment is in progress Auxiliary scale interval Lights uponly when theauxiliary scale interval is displayed 1 Bar graph C ct carat JJ 3 OZ oz ounce IB lb pound OZ fc o...
Страница 9: ...it function Function key Brief press Brief press Brief press Continuous press Longer continuous press Brief press Zero Tare key Brief press Brief press toggle switches the units to be displayed in suc...
Страница 10: ...the bubble rests in the center of the red circle on the level Position of air bubble in the level Caution Use caution when operating the adjusters on the square pan type to prevent them from lifting...
Страница 11: ...ng is loaded the display does not change after o n this case remove the load 0 In 3 Switching the measurement mode aaat Press the Function key Each time the key is pressed the unit of measurement chan...
Страница 12: ...te only the weight of the sample loaded in the tare Weighing only the weight of an added sample Press the I Zero Tare key The total weight of the sample loaded on the measurement pan is indicated as z...
Страница 13: ...nd or other vibrations Use the windshield or vibration dampers to protect against such adverse effects O0 0 DI Unstable Stable 4 When the read out indicator is reset to zero or the tare is subtracted...
Страница 14: ...cting the next item E SEL 0 Press the Function key The indication changes to the next item 2 5 E L D Limit function u 3 Selecting an item Pressing the Function key 3 RO l advances the function items a...
Страница 15: ...d 5 graduation is judged ranges 5 graduation or below including negative ranges are not judged The entire range is judged the entire range including the negative isjudged One point setup judges betwee...
Страница 16: ...when unsealed 4 3 Interface Section Displayed when Item Output Control Baud Rate Parity l h ia r l3 LI is set Set Value O C h L PR to 0t 1 a s s A n TV a i kU I t a or ci Description Stop output Outpu...
Страница 17: ...ts of Measurement Only when the balance is not sealed When values 8 1 5 u to 3 5 5 u are entered prior to use the desired unit of measurement to be displayed can be chosen simply by pressing the Funct...
Страница 18: ...ted simply pressing the Function key For more information on the units of measurement that can be set please refer to 4 2 Description of Functions on page 13 2 The units are displayed in the same sequ...
Страница 19: ...unit weight Total weight _ 500 g Sample unit weight 1 500 pcs Measured number of samples S lf samples to be counted deviate widely in weight or a higher measure of accuracy is desired it is recommend...
Страница 20: ...following states 1 The weight of one sample measurable unit weight is insufficient For the range of unit weights that can be measured and saved please refer to 11 Specifications on page 30 2 In the s...
Страница 21: ...ples O M ED 4 Saving the samples 20 pieces saved H D I Add approximately an equal number of samples as the number currently displayed Add them gradually in groups of approximately the same number unti...
Страница 22: ...ntage mode 0 u Press the Function key to disaplay 2 Setting the reference value I I i P 5Et I S Press the Set key The display begins flashing P 5 E t The balance is now ready for reference value setup...
Страница 23: ...weight of the reference sample is insufficient For the limit weight that can be saved limit weight please refer to 11 Specifications on page 30 2 While setting up the reference value in Step 2 the Se...
Страница 24: ...mit value 2 Numeric value setup method The limit value is entered with a key stroke JKThe limit values entered are held in memory even when the balance is powered down JKThe respective limit values fo...
Страница 25: ...ss the Zero Tare key to select the desired condition 5 Setting the judgement range ZE L EB L Q Press the Function key The display changes to Judgement Range Press the Zero Tare key to select the desir...
Страница 26: ...fter the lower limit value has been saved the balance displays it briefly and proceeds to the following setup JKIf One point setup was chosen the setup is complete 4 The upper limit value setup H SEt...
Страница 27: ...nction key The flashing moves to the digit on the immediate left Each time the key is pressed the flashing digit moves one position left When the leftmost digit is selected the flashing advances to th...
Страница 28: ...nuous pressing r j 11 1n 3B Upper limit value displayed Lower limit value displayed Measurement mode restored 2 If you make a mistake press the Function key during the setup of actual quantities or th...
Страница 29: ...with the balance installed perfectly level and without any load on the measurement pan 1 Activating the span adjustment function F u n c u Pressed continuously Key released Press and hold down the Fun...
Страница 30: ...pacity of the balance To implement a calibration as precisely as possible use a weight close to the capacity of the balance JS Calibration weights can be ordered from Shinko For ordering information p...
Страница 31: ...ent when the power is switched on Overcharging will also decrease battery life 2 When the balance is used for the first time after purchase the operating time may be shorter than when using a fully ch...
Страница 32: ...because the sample weight is insufficient Gross weight exceeded the capacity of the balance weight range container weight of sample A section of the mechanism is damaged A foreign object is caught bet...
Страница 33: ...1 100 12 2 Common Specifications 1 Weight measuring method Tuning fork vibration method 2 Tare subtraction range Total capacity 3 Liquid crystal display LCD Seven segments two segments in leading part...
Страница 34: ...1 1 3 0 0001 19 6 001 390 0 01 16 0 001 16 0 001 16 0 001 160 0 01 53 0 001 SJ 1200CE 1200 0 1 6000 1 42 0 01 2 6 o ooi 38 0 01 770 0 1 32 0 01 31 0 01 32 0 01 320 0 1 100 0 01 SJ 2200CE 2200 0 1 1100...
Страница 35: ...333 1 0 05 1 20337 1 21528 1 20565 0 12057 0 37500 penny weight 0 64301 0 12860 18 22917 291 66667 20 1 24 06741 24 30556 24 11306 2 41131 7 5 unit 1g 1ct 1oz 11b 1ozt 1dwt 1tl Hong Kong 1tl SGRMYS 1t...