Launch Test:

 

               Figure 5: Home View 

 

Figure 6:  File Explorer 

1.

 

 

Press the 

ON/OFF

 button to turn on the test set. 

2.

 

   If files are to be saved, press the 

FILE

 Button and tap the 

results

 folder.  Otherwise, proceed 

to step 4. 

3.

 

  

Tap the 

Create Directory

 soft key, enter a name for your new folder, and tap the 

Enter

 key.  

All trace files will be saved to this directory.  

4.

 

  Press the 

HOME

 button to display the Home Screen. 

5.

 

 

 

Tap the desired (Multimode or Single Mode) 

FTTA OTDR 

icon until it is selected and 

highlighted yellow. 

6.

 

   Press the 

SETUP

 button.  If an 

FTTA_RRU_Maintenance

 configuration file has already been 

loaded, proceed to step 9.    

7.

 

  

Tap the 

Load Config

 soft key.  Select 

FTTA_RRU_Maintenance.SM-OTDR

 for single mode 

fiber, or 

FTTA_RRU_Maintenance.MM-OTDR 

for multimode fiber.  

8.

 

  

Tap the 

Load as FTTA Config

 soft key. 

9.

 

  

Tap the 

Analysis

 soft key:   

 

Set 

BBU Jumper

 to 

Yes

 if there is a BBU Jumper Cable between the Launch Cable and OPP; 

Set 

BBU Jumper

 to 

No

 if the Launch Cable is directly connected to the OPP. 

 

Set 

RRU Jumper

 to 

Yes 

if there is a Junction Box/OPP at the top of the tower between the 

Trunk Cable and RRU; Set 

RRU Jumper

 to 

No

 if there is no Junction Box or no RRU Jumper. 

 

Leave other settings at default values. 

 

Tap the 

Analysis

 title bar to display the 

Link

 soft key. 

10.

 

 

Tap the 

Link

 soft key: 

 

 

Set Base

 Station ID 

to the Base Station or BBU Identifier. 

 

Set RRU

 ID 

to the RRU Identifier or sector (Alpha, Beta, Gamma). 

 

Set 

Fiber Code

 to the fiber number and polarity using the right ( 

) and left ( 

) arrow 

buttons. 

 

Set 

Change Fiber Nbr

 to 

Increment.

 

 

Set

 Distance Unit

 to 

Feet

 or 

Meters.

 

 

 

Содержание T-BERD/MTS-2000

Страница 1: ... fiber optic patch cord Launch Cable with connectors that match the OTDR Port and Fiber Under Test SC UPC SC APC LC UPC etc Optical Coupler to connect Launch Cable to BBU Jumper Cable or Trunk Cable The following information is required to complete the procedure Type of Fiber Multimode or Single Mode Type of Connectors SC UPC SC APC LC UPC etc Tower architecture o Is there a BBU jumper cable o Is ...

Страница 2: ...er 1 Inspect and clean the OTDR port on top of the test set 2 If the interface to the FUT is the BBU Jumper or Trunk Cable connect the cable to an optical coupler with the same connector type 3 Inspect and clean the FUT connected to the coupler or OPP 4 Inspect and clean fiber end face of the Launch Cable 5 Connect the Launch Cable to the OTDR port 6 Inspect and clean the other fiber end face of t...

Страница 3: ...nance SM OTDR for single mode fiber or FTTA_RRU_Maintenance MM OTDR for multimode fiber 8 Tap the Load as FTTA Config soft key 9 Tap the Analysis soft key Set BBU Jumper to Yes if there is a BBU Jumper Cable between the Launch Cable and OPP Set BBU Jumper to No if the Launch Cable is directly connected to the OPP Set RRU Jumper to Yes if there is a Junction Box OPP at the top of the tower between ...

Страница 4: ...loss or reflectance are shown in RED Use the right and left arrow button to Toggle between results for each wavelength Tap the Zoom soft key to enable zoom controls with the following buttons Toggle between automatic zoom and full trace view Expand and contract trace b SmartLink Summary view The FUT is displayed as a series of icons representing events connectors splices bends etc The center of th...

Отзывы: