OneAdvisor-1000 Portable Network Tester
OneAdvisor-1000 Portable Network Tester
Q U I C K C A R D
4
Figure 11: SmartLink View, Summary Results
R U N T E S T
1.
Press the Start soft key to start the test.
After auto-configuration, the OTDR will perform
a connection check to ensure that the
connection is
Good
. If the Connection is
Bad
,
disconnect the launch cable, and reconnect as
described on pages 1 and 2, cleaning every fiber
that fails the inspection test.
2.
The OTDR will perform an acquisition at the
configured wavelength.
Figure 12: SmartLink View, Event Results
Figure 10: Connection Check
1.
Tap the
SmartLink
tab at the left screen top to
select SmartLink view.
2.
Set the toggle switch at the right screen top to
Event Line
.
3.
View SmartLink results:
►
The FUT is displayed as a series of icons
representing events (front connector,
launch cable, connectors, splices, bends,
etc.) and distances. Swipe left or right to
view additional icons on spans with more
than 4 events.
►
The center of the display shows summary
results for the entire span for each
acquisition or wavelength. Tap any event
in the upper display to view detailed results
for that specific event.
►
Events are marked with a green check or
red
x
based on the Alarms setting.
Note:
the icon-based view may not be
available on older ONA-1000s when
acquisition
Mode = SmartAcq
. Rerun the test
with
Mode = Auto
or contact VIAVI to upgrade
your unit to add this feature.
V I E W S M A RT L I N K M A P