
VIAVI Solutions
Quick Card
T-BERD
®
/MTS-2000 Modular Test Set
DWDM Optical Channel Checker (OCC-4056)
The following procedure outlines how to connect to a fiber under test, configure OCC-4056 test setups,
run tests, and analyze results with a T-BERD/MTS-2000 equipped with an OCC-4056 module.
Note: The OCC-4056 module is used to monitor wavelengths on a live network. Care should
be exercised when unplugging fibers from the network as the fibers may be carrying traffic.
Equipment Requirements:
•
T-BERD/MTS-2000 equipped with Fiber Optics
Software Release V18.0 or greater
•
OCC-4056 DWDM OSA Module
•
Fiber optic cleaning and inspection tools
•
20-meter Fiber optic patch cord (Launch
Cable)
•
Optical Coupler to connect Launch Cable to
Fiber Under Test
Figure 1: Equipment Requirements
The following information is required to
complete the test:
•
Type of Connectors (SC UPC, SC APC, LC, etc.)
•
DWDM Wavelengths used
Figure 2: OCC-4056 Module
Fiber Inspection Guidelines:
•
Use the VIAVI P5000i or FiberChek Probe
microscope to inspect both sides of every
connection being used (OTDR Port, Launch
Cable, bulkhead connector, patch cord, etc.)
•
Focus the fiber on the screen. If dirty, clean
the connector.
•
If it appears clean, run inspection test.
•
If it fails, clean the fiber and re-run inspection
test. Repeat until it passes.
Figure 3: Inspect Before You Connect (IBYC)