Chapter 6
Performing Tests and Measurements
Channel Power Measurements
22130634
CX300 ComXpert Operation Manual
Rev. 001
February, 2022
6 - 19
3.
Press the meter Refresh button.
4.
Observe the SINAD meter and verify its reading is 12 dB or greater, and that the
meter bar color is green (indicates pass condition).
6.6 Channel Power Measurements
6.6.1 Scope of Test
This test is used to identify the following signal characteristics:
•
Total RF power
•
Power Ratio
•
Adjacent Channel Power Ratio (ACPR)
6.6.2 UUT Parameters
The example in this section assumes the following UUT parameters; adjust settings
according to the operational capabilities of the UUT.
6.6.3 Equipment Needed
The following equipment is required to perform the test procedures defined in this section:
•
CX300 ComXpert Test Set
•
RF Coaxial Cable (1 each)
Table 6-4 Spectrum Analyzer Channel Power - UUT Parameters
UUT Parameters
Transmit Frequency
Transmit Power
Transmit Modulation
Maximum Deviation
Maximum Modulation Frequency
Microphone/Mod Input
Modulation Input Level for Test Deviation
Squelch Tone Deviation
Squelch Tone Frequency
Transmit Frequency