Chapter 5
Test and Measurement Functions
CX300 Receive Functions
Rev. 001
CX300 ComXpert Operation Manual
22130634
5 - 24
February, 2022
5.5.1.1 Configuring the RF Receiver
To Configure the RF Receiver - Basic Setup
1.
Select
Communications Test
Mode of operation.
2.
Select measurement (e.g.
Analog
) and press the
Done
button
.
3.
Connect the incoming RF Signal to the RF Receiver Connector appropriate for
the characteristics of the incoming signal.
4.
Press the
Full Test Settings button
(
) to open the
Full Test Settings
menu
.
5.
Select the RF Receiver
Port
to which the RF signal is connected (step 3).
6.
Set the RF Receiver
Frequency
to the frequency of the incoming signal.
7.
Set the RF Receiver
Reference Level
to the power level of the generator/radio.
8.
Select an RF Receiver
IF Bandwidth
.
9.
Set
Level Control Mode
to
Auto
.
10.
Navigate to the
Analog Demod
settings. Select the
Demod Type
according to
the modulation of the incoming signal.
INPUT OVERLOAD
Do not overload input connectors.