Chapter 5
Test and Measurement Functions
CX300 Generators
Rev. 001
CX300 ComXpert Operation Manual
22130634
5 - 16
February, 2022
5.4.1.1 Configuring the RF Generator
The RF Generator should be configured according to the capabilities of the receiver. Refer
to the receiver’s input ratings to ensure the Generator Output Level does not exceed the
operational capabilities of the receiver.
The following information about configuring the RF Generator.
1.
Select
Communications Test
mode of operation.
2.
Select measurement mode (e.g.
Analog
) and press the
Done
button
.
3.
Press the
Full Test Settings button
(
) to open the
Full Test Settings
menu
.
4.
Select the RF Generator
Port
.
5.
Define the RF Generator
Frequency
.
6.
Define the RF Generator
Level
.
7.
If desired, define the RF Generator
Level Offset
.
8.
Set the RF Generator
Enable
toggle switch to
ON
.
The CX300 is now generating an RF signal.
Level
RF Generator
Level
should be set to a value appropriate for
the receive capabilities of the UUT. When configuring the
equipment for testing, use the RF Generator connector best
suited to the
Level
setting.
NOTE
The RF Generator
Level
should not exceed the approved
rating for the input port selected on the receiver.
Level Offset
The RF Generator
Level Offset
accounts for a loss or gain to
be inserted into the RF path between the CX300’s generator
output connector and the Unit Under Test. The
Level Offset
field defines the value that is applied to the RF Generator's
power level.
•
The offset value is indicated in +dB for positive (gain)
values.
•
The offset value is indicated in -dB for negative (loss)
values.
Table 5-7 RF Generator Controls and Settings (Continued)
Control/Setting
Description