756
JD700B Series User’s Guide
Figure 457 optical power measurement
NOTE
If the Wavelength is set to Auto, it displays “Auto” for the modulation frequency. It it is set
to Manual, the instrument displays a detected modulation frequency.
Reading a relative power
To view a relative power in dB, you can set a reference value automatically or manually.
Procedure
1
Press the
AMP/SCALE
hard key.
2
Toggle the
Display
soft key and select
Relative
.
3
Do one of the following:
To let the instrument set the last measurement as a reference, complete the following steps:
Toggle the
Relative Reference
soft key and select
Last
.
Press the
Reference Last
soft key.
NOTE
To set or release reference, you can also press and hold the physical round button on the
connected MP-series power meter for a few seconds until you hear a beep.
To set the reference manually, complete the following steps:
Toggle the
Relative Reference
soft key and select
Manual
.
Press the
Reference Manual
soft key.
Enter a value manually as you desire, and then press the
Enter
soft key.
Holding a measurement
Procedure
1
To hold the measurement, press the
HOLD
hard key.
Measurement is paused and the text “
Hold
” appears.
2
To resume the measurement, press the
HOLD
hard key again.
Measurement is resumed and the text “
Hold
” disappears.
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...