
Chapter 7
Using WCDMA/HSPA+ Signal Analyzer
JD700B Series User’s Guide
197
Measurement example
Figure 101 CDP Table measurement with WCDMA/HSPA+ signal analyzer
Setting display
If the CDP table spans over to the next page(s), you can change pages up and down to view your
measurement results.
Procedure
1
Press the
TRACE/DISPLAY
hard key.
2
Select
Page Up
or
Page Down
as needed.
Setting limit for modulation tests
Procedure
1
Press the
LIMIT
hot key.
2
Press the
Modulation Test Limits
soft key to set the limits.
3
Select the test item(s) and set the limit(s) depending on your selected measurement mode:
To set the limit for
Select
Set
Frequency error
Frequency Error
High Limit, Low Limit
Error vector magnitude
EVM
High Limit
Peak code domain error
PCDE
High Limit
Carrier feed through
Carrier Feed Through
High Limit
Correlated power in coded channel
CPICH
High Limit [Abs|Rel]
Low Limit [Abs|Rel]
Highest level among inactive channels
Max Inactive
High Limit
Relative code domain error
RCDE
High Limit (QPSK)
High Limit (16 QAM)
High Limit (64 QAM)
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...