T-BERD/MTS-8000 V2 - Quick Card “IL/ORL Referencing procedure” - Ref 78000A203/00 - March 2013 -  Copyright © 2013 JDSU 

2

1.  Inspect and clean connectors

Before connecting a 

fi

 ber into a test module, inspect and clean the module bulkhead 

and the 

fi

 ber jumper connectors.

!

Use the P5000i video inspection scope to verify the connector quality.

Use appropriate cleaning material (e.g. IBC™ cleaner, cotton swab, dust air 
sprays, etc...) and re-inspect to con

fi

 rm.

2.  IL/ORL referencing with the integrated FiberComplete® solution

1

Connect a 

fi

 ber jumper to the MTAU COM port of T-BERD/MTS 8000 V2 and one 

to the test port of the module of the T-BERD/MTS 6000A.

Refer to Fiber Characterization Getting Started Quick Card for proper module 
interconnections.

2

Press  

 to turn both units on and wait the completion of auto-test (45 sec.).

Activate the FCOMP function on both units

1

Press the hard key 

 and light the icon 

 with your 

fi

 nger or stylus.

2

Press the hard key 

 and touch the MTAU tab of the 8000 V2 unit. 

3

Press the soft key 

 until 

OUT A

 is selected on the 8000 V2 unit.

3.  Perform ORL and side by side IL Referencing

1

Touch the 

FCOMP

 tab 

.

2

Press the soft key 

 to enter the reference menu.

3

Press the soft key 

Take Refs

 

 in order to enter the referencing wizard.

4

Press 

Side/Side

 key 

 to start referencing process for 

ORL and side by side Insertion Loss .

5

When prompt, connect the test jumpers from the 8000V2 COM port to the 
mainframe power meter and from the 6000 test module to mainframe power 
meter, as shown below

T-BERD/MTS-8000 V2 - Quick Card “IL/ORL Referencing procedure” - Ref 78000A203/00 - March 2013 -  Copyright © 2013 JDSU 

3

6

Press OK to measure emitted power level.

7

When prompt, disconnect the test jumpers from the mainframe power meters and 
connect the non-re

fl

 ective terminator to the end of each test jumper, as shown 

below

8

Press OK to perform zero ORL referencing.