Operation, Remote Operation
RTCA/DO-260 Test Commands
Rev. L0
ATC-5000NG ATC/DME Test Set Operation Manual
1-2-5
February 2020
Subject to Export Control. See Cover Page for details.
Page 23
5.6.2
LOAD
5.6.3
NUMBER OF TRANSMISSIONS
5.6.4
REPETITION INTERVAL
5.6.5
RESET
5.6.6
SAVE
Command Syntax:
{:ATC |:ATC5000NG}{:DO260}{:LOAD}SP<filename>CR
Description:
This command loads a CSV scenario file (specified filename) from the internal storage area. A “*”
is returned upon completion of loading the file.
Example:
:ATC:DO260:LOAD test1.csv\r
Command Syntax:
{:ATC |:ATC5000NG}{:DO260}{:TIMING}{:TRANSMISSIONS |:TRANS}SP<numeric>CR
Description:
This command sets the number of transmissions.
Numeric:
{ 0, 20, 40, 60, 100, 200, 400, 600, 945, 1000, 2000, 4000, 6000, 10000} (decimal ASCII)
Example:
:ATC5000NG:DO260:TIMING:TRANSMISSIONS 20\r
Default:
0 (unlimited)
Query:
:ATC5000NG:DO260:TIMING:TRANSMISSIONS?\r
Return:
20
Command Syntax:
{:ATC |:ATC5000NG}{:DO260}{:TIMING}{:PERIOD |:PER}SP<numeric>CR
Description:
This command sets the transmission interval in ms.
Numeric:
10 to 2000 ms (decimal ASCII).
Example:
:ATC5000NG:DO260:TIMING:PERIOD 20\r
Default:
10 ms (100 repetitions per second)
Query:
:ATC5000NG:DO260:TIMING:PERIOD?\r
Return:
20
Command Syntax:
{:ATC |:ATC5000NG}{:DO260}{:RESET}CR
Description:
This command initializes the DO-260 test to the default values.
Default:
Attribute
Test Type
Trigger Source
Trigger Mode
Period (ms)
Number of Transmissions
Signal (All Generators)
Power (All Generators)
Phase (All Generators)
Path (All Generators)
Delay (All Generators)
Message Type
Value
Normal
GENA
Delay+
10
0 - Unlimited
OFF
-20 dBm
0 deg
Top
0 ns
None
Example:
:ATC:DO260:RESET\r
Command Syntax:
{:ATC |:ATC5000NG}{:DO260}{:SAVE}SP<filename>CR
Description:
This command saves the current scenario into the internal storage area with the specified
filename.
Example:
:ATC:DO260:SAVE test1.csv\r