RGS-2000NG TCAS Test Set Operation Manual
1-2-5
September 2021
Subject to Export Control. See Cover Page for details.
Page 33
5.8.1.13
Walk ON/OFF
5.8.2
LOAD
5.8.3
NUMBER OF TRANSMISSIONS
5.8.4
REPETITION INTERVAL
Command Syntax:
{:RGS | :RGS2000NG}{:DO260}{:TYPE}{:NORMAL } {:GENS}{:GENA |
:GENB | :GENC | :GEND | :GENE | :GENF }{:WALK}SP{ON | OFF}CR
Description:
This command enables or disables the walk status of the transmission using
the selected generator. By default, the walk status is enabled when the
trigger mode is set to walk. For the generator trigger source, the walk status
is disabled. If walk is disabled, the transmission begins the d elay value from
the trigger source. If walk is enabled, the transmission initially begins at the
delay value and then moves 50 ns every transmission until the end.
Default:
On
Example:
:RGS:DO260:TYPE:NORMAL:GENS:GENA:WALK OFF\r
Command Syntax:
{:RGS | :RGS2000NG}{:DO260}{:LOAD}SP<filename>CR
Description:
This command loads a CSV scenario file (specified filename) from the
internal storage area.
Return:
“*” is returned upon completion of loading the file.
Example:
:RGS:DO260:LOAD Test260.csv\r
Command Syntax:
{:RGS | :RGS2000NG}{:DO260}{:TIMING}{:TRANSMISSIONS |
:TRANS}SP<numeric>CR
Description:
This command sets the number of transmissions.
Numeric:
{0, 20, 40, 60, 100, 200, 400, 600, 945, 1000, 2000, 4000, 6000, 10000}
(decimal ASCII).
Setting to 0 means unlimited number of transmissions.
Default:
0 (unlimited)
Example:
:RGS:DO260:TIMING:TRANS 20\r
Command Syntax:
{:RGS | :RGS2000NG}{:DO260}{:TIMING}{:PERIOD | :PER}SP<numeric>CR
Description:
This command sets the transmission interval in ms.
Numeric:
10 to 2000 ms (decimal ASCII).
Default:
10 ms (100 repetitions per second).
Example:
:RGS:DO260:TIMING:PER 10\r