Falcon Service Manual
___________________________________
20
a) Release slightly the Temperature nut which is located on the right side of the Falcon. Take
care not to completely release the nut. In order to avoid scratching of the casing, hold the tool at
roughly 90 degrees to the nut, and try to avoid touching the casing as shown below.
c)
There are 3 black screws on each of the right and left sides of the Falcon. Use a standard
Philips screw driver to unscrew these screws as shown below, and place the screws aside to be
available for the assembly stage.
d)
c) Carefully lift the Top Cover of the Falcon, and place aside on a rigid surface where it is safe
from potential scratches.
5.2
Release the Tubes from the Pumps
In the Falcon/Pro the pumps are located in 2 rows of 5 or 6 pumps each on PVC0067 Falcon/Pro
Pump Sub-assembly Front and PVC0068 Falcon/Pro Pump Sub-assembly Back. In the
Falcon/Quad and Falcon/ABI+ there are 4 pumps on PVC0069 Falcon/Quad Pump Sub-assembly
Back. Each pump is wired on one end, and connected with a short silicon tube oriented towards
the middle of the system. The silicon tube is connected to the air exit of the pumps. Gently pull
off the tubes from each of the pump air exits as shown below.
VS0152, rev 1.3