V1410 / V1412 User Manual
95
Mainstream/Sidestream CO
2
Messages
Message
Description
SENSOR WARM_UP
EtCO2 module is connected to the monitor
and is ready to operate.
SAMPLE LINE DISCONNECT
Sample line is disconnected or disassembled from
the module during measurement.
CHECK SAMPLING LINE
Measurement numeric value (or pressure) has
exceeded the allowable range, or the sample line
is not connected properly.
NO BREATHS DETECTED
Respiration is not detected, or the sample line is
disconnected from the patient.
CHECK ADAPTER
Verify sampling line is not occluded.
ZERO IN PROGRESS
Zero calibration is in progress.
ZERO REQUIRED
Zero calibration is needed. Zero calibration is in
progress.
SENSOR FAULTY
Sensor (EtCO2 module) is not connected.
SENSOR DISCONNECT
Sample line is not connected to the sidestream
module.
MESUREMENT MODE
EtCO2 measurement is in progress.