Arrhythmia Guide
The arrhythmia guide is not species specific and is only a tool for comparison.
Name of Arrhythmia
Description
Tachycardia (TAC)
Heart rate is over 140 bpm for a minute.
Bradycardia (BRD)
Heart rate is under 40 bpm for a minute.
Premature Ventricular
Contraction (PVC)
Abnormal heart rate is over 1.
Ventricular Tachycardia (VTAC) Over 8 PVC is over 140 bpm.
Asystole (ASY)
No electric action of heart.
Bigeminy (BGM)
After normal heart rate once, two times or more PVC.
Trigeminy (TGM)
After normal heart rate at two times, two times or more PVC.
Ventricular (VENT)
Check the state of the ventricular.
Ventricular Fibrillation (VFIB)
A fibrillatory wave for 4 consecutive seconds.
Couplet (CPT)
Abnormal heart rate is two times.
Triplet (TPT)
Abnormal heart rate is more than three times.
Missed Beat (MIB)
ECG wave is missed intermediately.
R-on-T
PCV is detected before finishing T wave in the normal ECG
wave.
NOTE
The V1410 / V1412 provides 13 types of arrhythmia display in two alarm levels:
- High Alarm: VTAC, ASY
- Low Alarm: PVC, BGM, TGM, TAC, BRD, VENT, VFIB, CPT, TPT, MIB, R-on-T
If you press the
button during the arrhythmia alarm, the
icon is displayed on the
top of the screen and all alarms will be temporarily silenced for 2 minutes.
Although studied widely in human medicine, arrhythmia measurements have not been
validated in veterinary medicine and are offered only as an ancillary tool.