V1407 User Manual
77
Changing the IBP Label
1
To change the IBP label, press the Menu button one
time and select
LABEL
.
Mode
Setting
Range for Change
Description
SPEED
6.3, 12.5, 25, 50 mm/sec
Set the speed of IBP waveform.
LABEL
ABP, CVP, ICP, PAP
Set the level according to the
measurement site of BP. Change the alarm
setting range according to level.
UNIT
mmHg, kPa
Set the unit of IBP measurement.
* ZERO
ON
Adjust the zero point of the transducer.
SCALE
** OPTIMUM / FIXED
Set the wave scale.
TRACE
ON / OFF
Set the display of waveform.
* Adjust the zero point of the transducer: In order to obtain accurate results, make sure to adjust the zero
point of the transducer before measuring a patient.
** Fixed: Basic value according to the established label (ABP: 0–200, CVP: 0–50, ICP: 0–50, PAP: 0–50).
Optimum: Scale of waveform is automatically adjusted regardless of the level.