CAUTION
Respiration signals may be distorted due to radiated electromagnetic interference from an
ESU or other electromagnetic devices. Should this occur, try repositioning the ESU further
away or investigate with the manufacturer the possibility of utilizing a higher rated shielded
cable.
NOTE
Respiration is measured with the ECG and HR etc., using the ECG cable.
To set up apnea alarm, press the Menu button and select ALARM
RESP
APNEA.
- APNEA Setting Range: OFF, 10 – 40 seconds (every 5 seconds)
If an EtCO2 signal is detected by the monitor, the Respiration Rate will be determined
and displayed via the EtCO2 data as a priority over ECG respiration data.
Changing the Respiration Waveform
1
To change the gain of the respiration waveform, press the
Menu button one time, and select RESP
SIZE.
2
To change the speed of respiration waveform, press the Menu button one time and select
RESP
SPEED.
Mode settings are described in the following table:
Setting
Set Mode
Description
SIZE
X0.25, X0.5, X1, X2, X3, X4, AUTO
Set the size of RESP waveform.
SPEED
6.25, 12.5, 25, 50 mm/sec
Set the speed of RESP waveform.
TRACE
ON / OFF
Set the display of the waveform.