Vernier TRB-BTA Скачать руководство пользователя страница 1

1

6. Monitor the turbidity value.

Note

: Particles in the water will settle over time

and show a slow downward drift in turbidity readings; therefore, take your
readings soon after placing the cuvette in the sensor.

Videos

View videos related to this product at

www.vernier.com/trb-bta

Calibrating the Sensor

For many experiments, calibrating the Turbidity Sensor is not required.
However, periodic calibration will provide the most accurate results. To
calibrate:

1. Warm up the sensor by connecting it to the software for at least five minutes

prior to use.

2. Enter the calibration routine for your data-collection program.
3.

First Calibration Point

: Obtain the cuvette containing the Turbidity

Standard (100 NTU) and gently invert it four times to mix in any particles
that may have settled to the bottom.

Important

: Do not shake the standard.

Shaking will introduce tiny air bubbles that will affect turbidity readings.

4. Wipe the outside of the cuvette with a soft, lint-free cloth or tissue.
5. Holding the standard by the lid, place it in the Turbidity Sensor. Align the

mark on the cuvette with the mark on the Turbidity Sensor.

Important

:

These marks must be aligned whenever a reading is taken.

6. Close the lid.
7. Enter

100

as the value in NTU.

8. Remove the standard.
9.

Second Calibration point

: Prepare a

blank

by rinsing the empty cuvette

with distilled water, then filling it to the top of the line with distilled water.

Important

: The bottom of the meniscus should be at the top of the line for

every measurement throughout this test. This volume level is critical to
obtain correct turbidity values.

10. Screw the lid on the cuvette. Wipe the outside with a soft, lint-free cloth or

tissue.

11. Holding the cuvette by the lid, place it into the slot of the Turbidity Sensor.

Makec sure that the marks are aligned. Close the lid.

12. Enter

0

as the value in NTU. You are now ready to collect turbidity data.

13. Store the calibration by following the directions under the Storage tab

(Logger

Pro

and LabQuest only).

14. You are now ready to collect turbidity data.

Specific instructions for each software platform can be found here:.

l

For instructions on calibration using Logger 

Pro

computer software, see

www.vernier.com/til/2341

l

For instructions on calibration using LabQuest App, see

www.vernier.com/til/3394

l

For instructions on calibration using Graphical Analysis with a
Chromebook, see

www.vernier.com/til/3631

Turbidity Sensor

(Order Code TRB-BTA)

Turbidity is a measure of water’s lack of clarity and is
an important indicator of water quality. Water with
high turbidity is cloudy, while water with low
turbidity is clear. The cloudiness is produced by light
reflecting off particles in the water; therefore, the more particles in the water, the
higher the turbidity.

Note:

Vernier products are designed for educational use. Our products are not

designed nor are they recommended for any industrial, medical, or commercial
process such as life support, patient diagnosis, control of a manufacturing
process, or industrial testing of any kind.

What's Included

l

Turbidity Sensor

l

Turbidity Accessories Kit (includes one empty cuvette and one cuvette
containing 100 NTU StablCal

®

Formazin Standard)

l

Safety Data Sheet for StablCal Formazin Standard

Compatible Software

See

www.vernier.com/manuals/trb-bta

for a list of interfaces and software

compatible with the Turbidity Sensor.

Getting Started

1. Connect the sensor to the interface (LabQuest Mini, LabQuest 2, etc.).
2. Start the appropriate data-collection software (Logger

Pro

, Logger Lite,

LabQuest App) if not already running, and choose New from File menu.

See the following link for additional connection information:

www.vernier.com/start/trb-bta

Using the Product

Collecting Data

1. Warm up the sensor by connecting it to the software for at least five minutes

prior to use.

2. Gently invert the sample water to mix in any particles that may have settled

to the bottom.

Important

: Do not shake the sample. Shaking will introduce

tiny air bubbles that will affect turbidity.

3. Empty the distilled water from the cuvette and rinse it with sample water.

Fill the cuvette to the top of the line with sample water.

4. Screw the lid on the cuvette. Wipe the outside with a soft, lint-free cloth or

tissue.

5. Hold the cuvette by the lid and place it into the Turbidity Sensor. Make sure

the marks are aligned. Close the lid.