Vermason ZVM LIMMIT COMPARATOR Скачать руководство пользователя страница 1

UNIT C, 4TH DIMENSION, FOURTH AVENUE, LETCHWORTH, HERTS, SG6 2TD UK

Phone: 0044 (0) 1462 672005, • E-mail: 

[email protected]

, Internet: 

Vermason.co.uk

TB-7564

 Page 1 of 2

© 2014 Vermason

ZVM LIMIT COMPARATOR

Operating Instructions

1.  The purpose of the ZVM Limit Comparator is to verify the 

     calibration of the 222755 Zero Volt Monitor by checking 

     four operator conditions: FAIL LOW, PASS (low limit), 

     PASS (high limit), and FAIL HIGH.

2.  There are 10 resistor settings on the Limit Comparator,

  which can be set using the rotary switch.  Positions 3 

  through 8 have installed values.  Positions 1, 2, 9, and 10 

  can be installed by the user for one set each of custom 

  low and high limits.  The standard settings are as follows:

POSITION:

 

1.) optional low limit fail low

 

2.) optional low limit pass

 

     ------------------------------

 

3.) 1.91 Megohms FAIL LOW

 

4.) 1.91 Megohms PASS (ZVM default settings)

 

5.) 10 Megohms PASS

 

6.) 10 Megohms FAIL HIGH

 

     ------------------------------

 

7.) 35 Megohms PASS

 

8.) 35 Megohms FAIL HIGH

 

9.) optional high limit pass safe

 

10.) optional high limit fail high

PLEASE NOTE THAT THE ZERO VOLT MONITOR 

OPERATOR LIMIT COMPARATOR IS USED TO 

CHECK ONE OPERATOR AT A TIME

3.  Ensure that the Zero Volt Monitor to be checked is 

     set up as described in the operating instructions and 

     has power.
4.  Connect the plug from the limit comparator into the 

     OPERATOR 1 remote jack. 

January 2014

Figure 1.  Vermason 

222770

 ZVM Limit Comparator

Made in the

United States of America

  TECHNICAL BULLETIN TB-7564

5.  Turn the rotary knob on the Limit Comparator to “x

  Ohms LOW” (select  x = 1.91 Megohms or the

  optional low limit fail low, according to what operator

  low limit the Zero Volt Monitor is calibrated).  

  Observe the operator 1 LED display on the front of 

  the Zero Volt Monitor.  The yellow LED should be on 

  and the audible alarm should sound, indicating the 

  fail low condition.
6.  Turn the rotary knob on the Limit Comparator to “x

  Ohms PASS” (select x = 1.91 Megohms or the

  optional low limit pass safe, according to what 

  operator low limit the Zero Volt Monitor is calibrated). 

  The green LED on the Zero Volt Monitor operator 1 

  display should be on, indicating the low limit pass 

  safe condition.
7.  Turn the rotary knob on the Limit Comparator to “x

  Ohms PASS” (select x = 10 Megohms, 35 Megohms 

  or the optional high limit pass safe, according to 

  what operator high limit the Zero Volt Monitor is 

     calibrated). The green LED on the Zero Volt Monitor 

     operator 1 display should be on, indicating the high 

     limit pass condition.
8.  Turn the rotary knob on the Limit Comparator to “x 

  ohms HIGH”  (select x = 10 Megohms, 35 Megohms 

  or optional high limit fail high, according to what 

     operator high limit the Zero Volt Monitor or is 

     calibrated). The red LED on the Zero Volt Monitor 

     operator 1 display should be on and the audible 

     alarm should sound, indicating the fail high condition.
9.  Disconnect the Limit Comparator plug from the

  operator 1 remote jack.  Plug the Limit Comparator 

  plug into the operator 2 remote jack and repeat the 

  four tests in steps 5 through 8 to test operator 2.
10. The correct color LEDs must light for each step for 

      the Zero Volt Monitor to completely pass the 

      calibration check.  If an incorrect LED comes on 

      during any portion of the test (example:  expecting 

      the green LED to light in step 7 but the Zero Volt 

      Monitor red LED is on) recalibrate the Zero Volt 

      Monitor and test each operator again with the Zero 

      Volt Monitor cal limit comparator.

INSTALLING OPTIONAL LOW AND HIGH LIMIT 

RESISTANCES

A.)  Use a hex wrench to remove the rotary switch 

      knob. Unscrew the 2 screws on the back of the unit 

      and disassemble the unit.  
B.)  Two resistances must be installed for each 

      limit:  one for PASS and another for FAIL. These 

      two resistances should be ±10% of the calibrating 

      resistance.  For example:

Отзывы: