10.0 Fuses and Power Settings
10.0
Fuses and Power Settings
10.1
Fuses
F1: Refer to the circuit diagrams.
F1
Internal
Supply
Fuses
Fitted if supply cable fitted.
Fitted on board to some types
of EMC filter.
GEC Safeclip of the type shown
(glass type F up to 16 A)
38 mm x 10 mm type F fitted on
EMC filter circuit board(s)
Model
Volts
Phases
Supply Fuse Rating
SNF
220-240
1-phase
4 A
SNF
110-120
1-phase
10 A
10.2
Power Settings
The power limit settings (parameter OP.Hi) for this model are voltage dependant. The
figures represent the maximum percentage of time that controlled power is supplied to
the elements. Do not attempt to “improve performance” by setting a value higher than
the recommended values. To adjust the parameter refer to the "Changing the
Maximum Output Power" of the control section of the manual.
Volts (V)
200
208
220
230
240
Power (%)
100
100
100
100
100
Please refer to the rating label for product specific information.