6.0
Maintenance
6.1
General Maintenance
Preventive rather than reactive maintenance is recommended. The type and frequency
depends on the product use; the following are recommended.
6.2
Maintenance Schedule
CUSTOMER
QUALIFIED PERSONNEL
DANGER! ELECTRIC SHOCK
. Risk of fatal injury. Only electrically
qualified personnel should attempt these maintenance procedures.
Maintenance
Procedure
Method
Frequency
Daily Weekly Monthly
Bi-
Annually
Annually
Safety
Over-Temperature Safety
Circuit
(if fitted)
Set an over-temperature setpoint lower than the
displayed temperature and check for an over-tem-
perature alarm as detailed in this manual
Over-Temperature Safety
Circuit
(if fitted)
Electrical measurement
Door Seal
Visual inspection, checking for splits and fraying
Door Seal
Replace where necessary
Chimney / Extraction
Check and clean if necessary
Electrical Safety (external)
Visual check of external cables and plugs
Electrical Safety (internal)
Physically check all connections and cleaning of
the power plate area
Function
Temperature Calibration
Tested using certified equipment, frequency
dependent on the standard required
Operational Check
Check that all functions are working normally
Operational Check
Thorough inspection and report incorporating a
test of all functions
Performance
Element Circuit
Electrical measurement
Power Consumption
Measure the current drawn on each phase / cir-
cuit
15