![VERDER CARBOLITE GERO E Series Скачать руководство пользователя страница 55](http://html.mh-extra.com/html/verder/carbolite-gero-e-series/carbolite-gero-e-series_installation-operation-and-maintenance-instructions_1003652055.webp)
13.0
Fuses and Power Settings
13.1
Fuses
F1 - F2: Refer to the circuit diagrams.
F1
Internal
Supply
Fuses
Fitted if supply cable fitted.
Fitted on board to some types
of EMC filter.
GEC Safeclip of the type shown
(glass type F up to 16 A)
38 mm x 10 mm type F fitted on
EMC filter circuit board(s)
F2
Auxiliary
Circuit Fuses
Fitted on board to some types
of EMC filter.
May be omitted up to 25 Amp/
phase supply rating.
2 Amps glass type F
On board: 20 mm x 5 mm
Other: 32 mm x 6 mm
Customer
Fuses
Required if no supply cable
fitted.
Recommended if cable fitted.
See rating label for current;
See table below for fuse rating.
Model
Phases
Volts
Supply Fuse Rating (Amps)
F1
EHA 12/300B
1-phase
100-120
16 A
EHA 12/300B
1-phase
200-208
10 A
EHA 12/300B
1-phase
220-240
10 A
Model
Phases
Volts
Supply Fuse Rating (Amps)
F1
55
13.0 Fuses and Power Settings
Содержание CARBOLITE GERO E Series
Страница 12: ...2 0 Installation Figure 2 3 4 Figure 2 3 5 Figure 2 3 6 Figure 2 3 7 Figure 2 3 8 12 ...
Страница 25: ...3 6 Navigation Diagram 25 3 0 2416 Controller ...
Страница 38: ...6 0 Operation than inert gases such as nitrogen 38 ...
Страница 52: ...11 0 Wiring Diagrams 11 3 WA 12 30 52 ...
Страница 59: ...Notes Service Record Engineer Name Date Record of Work ...