7.0
Maintenance
7.1
General Maintenance
Preventive rather than reactive maintenance is recommended. The type and frequency
depends on the product use; the following are recommended.
7.2
Maintenance Schedule
CUSTOMER
QUALIFIED PERSONNEL
DANGER! ELECTRIC SHOCK
. Risk of fatal injury. Only electrically
qualified personnel should attempt these maintenance procedures.
Maintenance
Procedure
Method
Frequency
Daily Weekly Monthly
Bi-
Annually
Annually
Safety
Safety Switch Function
Set a safe temperature above ambient, and open
the door to see if the heater light goes out
Safety Switch Function
Electrical measurement
Element Over-Temperature
Circuit
Adjust the set-point on the element over-tem-
perature controller to a lower value than the cur-
rent displayed temperature on the main
controller and ensure the heater lights go out.
Element Over-Temperature
Circuit
Electrical measurement
Over-Temperature Safety
Circuit
(if fitted)
Set an over-temperature setpoint lower than the
displayed temperature and check for an over-tem-
perature alarm as detailed in this manual
Over-Temperature Safety
Circuit
(if fitted)
Electrical measurement
Door Plug
Visual inspection, checking the seal and whether
it is free of damage
Door Plug
Replacement where necessary
Chimney / Extraction
Check and clean if necessary. Empty the col-
lection tray at the rear
Electrical Safety (external)
Visual check of external cables and plugs
Electrical Safety (internal)
Physically check all connections and cleaning of
the power plate area