The constituent elements of the PID control are also shown on the screen as
Proportional, Integral and Derivative (see Fig 53.).
10.2
Heater On / Off Control
The heater power can be switched on or off at any time using the red and blue buttons
on the right hand side of the screen (see Fig 54.). On system power up the on/ off
status of the heater power is determined by an ‘Initial Heater Setting’ which is located
on the General Settings screen (see Fig 54.).
General Settings Screen is accessed via:
Home Screen
→
Main Menu
→
System Configuration
→
General Settings
The factory default for this setting is ‘off’. This ensures the furnace will not heat up after
applying power for the first time during installation and hence protects any items that
may still be packed inside the main chamber.
Fig 54 - Screen-view: General Settings
49
10.0 Temperature Control