7.0 Running a Test
7.0
Running a Test
7.1
Set up
Before running the test for the first time, ensure the tray lift calculation has been
performed as described in section 8.0.
Before every test, the weights of the sample tray and sample must be determined as
described in section 7.13.
7.2
Selecting a Recipe
To start a test a recipe must first be created and then selected. Refer to the section 14.0
for details on creating a recipe. To select a recipe go to the Recipe Library by pressing
the ‘View Recipe Library’ button on the Main Menu. Select the required recipe from the
displayed list and then press the ‘Run’ button (see section 13.5). The ‘Home’ screen will
then be displayed with the selected recipe in the recipe name field, the temperature
setpoint from the recipe and depending on the recipe type, either the calibration factor
or binder % from the recipe will be displayed as shown in Fig 36. The furnace will heat
the main chamber to the temperature specified in the recipe setpoint. If the furnace has
just been switched on from cold, it should be allowed to dwell at the setpoint
temperature for 30 minutes. The after burner will be heated to 900 °C. Tare balance
(see section 17.2), then load the sample and tray. The operator should check that the
pre-requisite test conditions are satisfied before attempting to start the test, refer to the
section 7.3. If the conditions are satisfied, the test can be started by pressing the ‘Run
Test’ button.
Note: the text on this button will change, depending on the state of the system, as
shown in the table below.
Fig 38 - Screen-view: Home Screen Recipe Settings
32
Содержание CARBOLITE GERO ABA 7/35B
Страница 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Страница 13: ...Fig 7 Front brick cover Fig 8 Calibration Plate Fig 9 Functional parts of the ABA 13 2 0 Supplied Item List ...
Страница 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Страница 40: ...7 0 Running a Test Fig 46 Screen view Weight Entry Number Pad Fig 47 Screen view Successful Weight Entry 40 ...
Страница 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Страница 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Страница 93: ...Notes Service Record Engineer Name Date Record of Work ...