background image

Meditell 2 User Manual VI49.2 

 

- 6 -

 

3.0 Indicators and Sounders 

 

3.1  ECU Green Power LED 

The Green Power LED shows the status of the 

Meditell 2

 power supply. 

 
LED Illuminated  

 

 

 

 

- Mains Supply Present. 

LED flashes once every 5 seconds 

 

 

- Mains supply off and operating on battery backup.  

LED Flashers once every 20 seconds 

 

 

- Battery Save mode, battery flattened. 

LED Off  

 

 

 

 

 

- No power to the System. 

 

3.2  ECU Amber Fault LED 

The amber Fault LED illuminates to indicate the presence of a fault conditions. 
 
LED Off  

 

 

 

 

 

- System Normal. 

LED Illuminated  

 

 

 

 

- System Wiring Fault. 

LED flashing once per second.   

 

 

- System Wiring Fault 

– Short circuit. 

LED flashes once every 5 seconds 

 

 

- Mains supply off and operating on battery backup       

LED Flashers once every 20 seconds 

 

 

- Battery Save mode, battery flattened. 

 

3.3  ECU Red Call LED 

The Red “Call” LED indicates the status of any “Call” Alarm activations present on the Meditell 2 system. 
 
LED Off  

 

 

 

 

 

-  No Call Alarm Present. 

LED flashing once per second.   

 

 

-  

“Standard Call” Activated. 

LED flashes Twice per second   

 

 

-  

“Priority Call” Activated.

 

 

3.4  ECU Internal Buzzer 

 

Buzzer Off 

 

 

 

 

 

-  No Call Alarm Present. 

Buzzer Sounding Once per second. 

 

 

-  

“Standard Call” Activated. 

Buzzer Sounding Twice per second 

 

 

-  

“Priority Call” Activated. 

Buzzer Sounding once every 30 seconds 

 

- Battery Save mode, battery flattened.

 

 

3.5  Field Warning Devices 

 

Indicators / Sounders Off 

 

 

 

-  No Call Alarm Present. 

Indicators / Sounders activated once per second. 

-  

“Standard Call” Activated. 

Indicators / Sounders activated Twice per second 

-  

“Priority Call” Activated.

 

 

 

Содержание Meditell 2

Страница 1: ...Meditell 2 Disabled Persons Response Alarm User Operation Guide Document VI49 2 Protecting People www ventcroft co uk...

Страница 2: ...s trademarks of their respective companies Notice of Liability Every effort has been made to ensure that this manual instructions contains accurate and current information However Ventcroft Ltd shall...

Страница 3: ...nerate a Call Alarm Using a Ceiling Pull Cord Device 7 4 2 Mute a Call Alarm at the ECU 7 4 3 Reset a Call Alarm using a Remote Reset Device 7 4 4 Reset a Call Alarm using ECU 7 4 5 System Test Reset...

Страница 4: ...they produce a more urgent alarm to summon other carers Reset Devices which are generally wall mounted and consist of push button devices which are pressed to cancel reset an alarm activation Reset d...

Страница 5: ...low the Meditell 2 system to continue to operate During Battery back Up Mode as in Standby by mode the ECU is waiting to receive a call The Green Power LED Amber Fault LED will flash once every 5 seco...

Страница 6: ...ins supply off and operating on battery backup LED Flashers once every 20 seconds Battery Save mode battery flattened 3 3 ECU Red Call LED The Red Call LED indicates the status of any Call Alarm activ...

Страница 7: ...ll continue to in alarm 4 The ECU internal Buzzer will continue to briefly bleep once every 10 seconds until the alarm is reset 4 3 Reset a Call Alarm using a Remote Reset Device 1 Locate a Remote Res...

Страница 8: ...ll sound and flash twice per second The Indicator Mounted on the Call Device will illuminate in sympathy with the remote warning devices 4 8 Reset a Priority Call using a Remote Master Reset Device 1...

Страница 9: ...esetting when required The ECU Front mounted Reset Mute button can be used to perform a simple self test of the system warning devices See Section 3 5 for details The times mentioned above should be c...

Страница 10: ...Meditell 2 User Manual VI49 2 10 5 3 Routine Test Results TESTED DATE TESTED BY COMMENTS...

Страница 11: ...Meditell 2 User Manual VI49 2 11 NOTES...

Страница 12: ...uct please do not hesitate to ring Technical Help 44 0 1928 581098 Ventcroft Ltd Goddard Road Astmoor Industrial Estate Runcorn Cheshire WA7 1NQ United Kingdom Tel 44 0 1928 581098 Fax 44 0 1028 58109...

Отзывы: