background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

VR2Tx Quick Guide 

Place VR2Tx on clean, stable surface so hydrophone hangs off edge.  We 
recommend working in a controlled environment. 

Insert one steel rod into the Attachment hole and other one in the Bluetooth 
Activator Hole on the end-cap and hold on to the rods.  

Apply pressure to the rod in the Bluetooth Activator Hole to turn end-cap 
counter-clockwise when you are facing the hydrophone.  Remove rod when 
end-cap turns easily and continue turning by hand.  

Slide end-cap and cylinder apart until the battery  
compartment has cleared the cylinder.  Be very  
careful not to damage the threads on the end-cap  
or the VR2Tx 

won’t close and seal properly. 

Open Case 

Electronics inside the case 
must 

NEVER come in 

contact with any water 

or 

the VR2Tx will be damaged. 

The battery is held in the battery compartment by a plastic pin spanning the opening. Remove 
the thin O-ring before removing the pin (O-ring holds pin in place). Downward force on battery 
may be needed to allow pin to move. 

Battery 

The battery is connected to VR2Tx on the side of the case. When 
installing the battery, listen for a “click” and watch that light flashes 
four times and then twice quickly every 5 seconds. When 
disconnecting battery, press the release tab on the  
battery connector and then  
pull connector halves apart. 

This VR2Tx Quick Guide provides information on key VR2Tx 
procedures.  Full details for 

all

 VR2Tx procedures are found 

in the VR2Tx User manual and at www.vemco.com.  We 
strongly recommend you study the full user manual before 
using this quick guide. 

Make sure the two O-rings in the end-cap, and the 
corresponding O-ring surface inside the Outer 
Cylinder, are clean and debris-free. If the VR2Tx 
has previously  
been deployed,  
replace the  
O-rings. 

O-rings 

Doing so may damage the receiver and render 
it unable to detect transmitters. 

Do not bump the VR2Tx 

hydrophone or expose it to 

extreme temperature variations. 

TIP:

 Write the date on the battery. 

Close Case 

Close the VR2Tx case by doing the reverse of opening it, but 
first 

put a new desiccant pack in the bottom of the tube

Slide the battery end into the Outer Cylinder. Rotate the end-
cap clockwise, using  
the steel bars when  
needed.  

Stop turning when the 
gap between the two 
halves is closed.  
Over tightening could 
damage the case. 

A desiccant pack helps 

reduce the occurrence of 

condensation in the VR2Tx 