Figure 6-1 A Single Return
6.2.2 Multiple Returns
Multiple laser returns are possible from any single laser shot because of beam divergence. When a laser pulse leaves the
sensor it slowly, gradually grows larger. A pulse can be big enough to hit multiple objects producing multiple reflections.
Usually, the farther away a reflection starts, the weaker it is at the detector. Bright or retroreflective surfaces may flip that,
however.
The VLP-16 analyzes multiple returns and reports either the strongest return, the last return, or both returns, depending
on the laser return mode setting. If the setting is Strongest and a pulse produces multiple returns, only the strongest reflec-
tion results in a measurement. Likewise, if the setting is Last, only the last (time-wise) reflection results in a measurement.
This could be used to measure distances to the ground from the air.
6.2.3 Dual Return Mode
shows the majority of a laser pulse striking the near wall while the remainder hits the far wall.
The Dual return mode setting allows you obtain both measurements.
Note that the sensor records both returns only when the separation between the two objects is one meter or more.
Chapter 6 • Key Features
33
Содержание VLP-16
Страница 1: ...VLP 16 User Manual 63 9243 Rev D ...
Страница 64: ...Figure 9 9 Single Return Mode Timing Offsets in µs 64 VLP 16 User Manual ...
Страница 86: ...http 192 168 1 201 cgi setting laser on 204 OK Sensor laser is On motor rpm is 301 86 VLP 16 User Manual ...
Страница 106: ...C 6 Puck Hi Res Optical Drawing Figure C 6 Puck Hi Res OpticalDrawing 86 0129 Rev A 106 VLP 16 User Manual ...
Страница 109: ...D 1 Interface Box Wiring Diagram Figure D 1 Interface Box Wiring Diagram 86 0107A Appendix D Wiring Diagrams 109 ...
Страница 110: ...D 2 Interface Box Schematic Figure D 2 Interface Box Schematic 69 8230A 110 VLP 16 User Manual ...