PersonalScope
HPS10
Technical Doc
18/11/2002
©
Страница 1: ...PersonalScope PersonalScope PersonalScope PersonalScope HPS10 Technical Doc 18 11 2002...
Страница 2: ...2 Velleman HPS10 Technical Doc IN THE PACKAGE HPS10 unit User Manual PROBE60S OPTIONAL CASEHPS hard case BAGHPS special bag for HPS10 HPS40 GIB general instruments bag soft PS905 9Vdc 500mA adaptor...
Страница 3: ...VAL Prepared by Approved by Customer acceptance by Please return signed copy DATE DATE DATE Includes HPS10 unit Manual and documents Safety measuring probe PROBE60S Model type Description Sample numbe...
Страница 4: ...ve and For the manufacturer Date 03 05 2002 Signature _____________________________ Name Stephan Santens R D manager EN 55022 Limits and methods of measurement of radio interference characteristics of...
Страница 5: ...Velleman 5 HPS10 Technical Doc VELLEMAN Instruments HPS10 1 CH Personal Scope EN55022 Field strenght Start frequency 30MHZ End frequency 1GHz 30dB Probe correction attenuation...
Страница 6: ...6 Velleman HPS10 Technical Doc VELLEMAN Instruments HPS10 1 CH Personal Scope EN55022 Field strenght Start frequency 30MHZ End frequency 1GHz 30dB Probe correction attenuation...
Страница 7: ...e D10 D11 PRLL4001 rectifier diode type D6 D7 D13 BAS45AL Low leakage diode type D8 BAV70 High speed double diode type D9 2A Pico fuse F1 LM2940S5 0 3 1A low drop regulator IC1 MC14052B analog multipl...
Страница 8: ...7 PLCC44 PLCC socket for processor IC7 96120205 Header connector female 20P For LCD connection HDR1X20 Header connector male 20P For LCD connection PG12864 Graphic LCD LCD1 3mm green led 3mm green LD1...
Страница 9: ...1K R27 1K GND PWM L2 220 L3 220 C24 100 AGND V2 C2 100 D1 BAT54 GND C3 100 C4 100 GND D2 BAT54 C5 100 GND R32 1K GND C25 100 L4 220 C26 100 AGND V2 V1 C27 100 T1 BDP32 T2 BDP31 AGND GND ZD1 BZD27 C10V...
Страница 10: ...1K V 6 R29 1K V2 C8 100n C9 100n R46 33K R47 33K ATT5 ATT4 R50 100K C38 1n5 DG ND DG ND R18 10K R19 10K ATT1 V 1 C39 18p AG ND RV1 47K OF FSET C10 100n AG ND V 2 R36 4K7 C40 27p RV2 100 GAIN V2 C11 1...
Страница 11: ...IN2 5 IN2 6 OUT1 1 OUT2 7 IC8 LM 358M 8 V2 V2 YPOS AGND DGND R12 47K R13 47K C33 470n C34 470n R14 270K DGND DB 0 7 DB 1 8 DB 2 9 DB 3 10 DB 4 11 DB 5 12 DB 6 13 DB 7 14 E 6 R W 5 D I 4 CS1 15 CS2 16...
Страница 12: ...12 Velleman HPS10 Technical Doc...
Страница 13: ...t Version with arrow keys and short press setup key to mark it then leave setup by long press ing setup key You should see the Version screen Now press ALL the keys one after the other starting with t...
Страница 14: ...Doc STEP 1 Offset adjustment Adjust RV1 until pass is indicated The trace on the screen must then remain stable See trimmer location TRIMMER LOCATION TIP Pressing Right or Left arrow key scrolls thro...
Страница 15: ...obe to the scope probe test output If necessary adjust CV1 until pass is displayed See trimmer location Press RIGHT ARROW KEY STEP 3 High level transient Connect the x1 probe to the scope probe test o...
Страница 16: ...k cir cuit round relay RY1 T7 Press RIGHT ARROW KEY STEP 6 0 1V div Gain calibration Connect the x1 probe with a 0 17Vdc 170mV supply If necessary adjust RV2 until pass is displayed Press RIGHT ARROW...
Страница 17: ...stment if fail check cir cuit round IC2 IC3 IC4 Press RIGHT ARROW KEY STEP 8 0 4V div test Connect the x1 probe with a 3 4Vdc supply No adjustment if fail check cir cuit round IC2 IC3 IC4 Press RIGHT...
Страница 18: ...round IC2 IC3 IC4 Press RIGHT ARROW KEY STEP 11 Trace center Adjustment Disconnect the test probe Adjust RV4 until pass is indi cated Press RIGHT ARROW KEY STEP 12 Calibration END If all steps where...
Страница 19: ...ith X10 probe 0 5dB accuracy dBV measurements 0dBV 1V From 75dB tot 38dB up to 58dB with X10 probe 0 5dB accuracy True rms measurement From 0 1mV to 80V up to 400Vrms with X10 probe 2 5 accuracy Peak...
Страница 20: ...A division of Velleman Components Legen Heirweg 33 9890 Gavere BELGIUM Europe www vellemaninstruments com HPS10TD1 0...