9 Adjustment menu
VEGAPULS C 11 • Two-wire 4 … 20 mA
58340-EN-200806
•
Further recommendations
– False signal suppression via adjustment app or PACTware/DTM
Overflow basin
•
Vessel
– Large volume
– Partly installed underground
•
Process/measurement conditions:
– Partly strongly agitated surface
– Multiple reflections through flat vessel ceiling
– Condensation, dirt deposits on the sensor
– Flooding of the sensor antenna
Vessel/Collecting basin
•
Vessel:
– Large volume
– Upright cylindrical or rectangular
•
Process/measurement conditions:
– Slow filling and emptying
– Smooth medium surface
– Condensation
Plastic tank (measurement through the vessel top)
•
Process/measurement conditions:
– Measurement through the tank top, if appropriate to the applica-
tion
– Condensation on the plastic ceiling
– In outdoor facilities, water and snow on vessel top possible
•
Further recommendations
– With measurement through the tank top false signal suppres-
sion via adjustment app or PACTware/DTM
– When measuring through the tank top in outdoor areas protec-
tive roof for the measuring point
Transportable plastic tank (IBC)
•
Process/measurement conditions:
– Material and thickness different
– Measurement through the vessel top, if appropriate to the
application
– Changed reflection conditions as well as jumps in measured
values when changing vessels
•
Further recommendations
– With measurement through the tank top false signal suppres-
sion via adjustment app or PACTware/DTM
– When measuring through the tank top in outdoor areas protec-
tive roof for the measuring point
Gauge measurement in waters
•
Process/measurement conditions:
– Slow gauge change
– Extreme damping of output signal in case of wave generation
– Ice and condensation on the antenna possible
– Floating debris sporadically on the water surface