background image

6

2 Product description

VEGAPULS 66 • Profibus PA

47140-EN-210621

2  Product description

2.1  Configuration

The type label contains the most important data for identification and 

use of the instrument:

2

1

15

13

14

12

11

5

3

6

4

7

8

10

9

Fig. 1: Layout of the type label (example)
1  Instrument type

2  Product code

3 Approvals

4  Power supply and signal output, electronics

5  Protection rating

6  Measuring range

7  Process and ambient temperature, process pressure

8  Material wetted parts

9  Hardware and software version

10  Order number

11  Serial number of the instrument

12  Data matrix code for VEGA Tools app

13  Symbol of the device protection class

14  ID numbers, instrument documentation

15  Reminder to observe the instrument documentation

The type label contains the serial number of the instrument. With it 

you can find the following instrument data on our homepage:

• 

Product code (HTML)

• 

Delivery date (HTML)

• 

Order-specific instrument features (HTML)

• 

Operating instructions and quick setup guide at the time of ship-

ment (PDF)

• 

Order-specific sensor data for an electronics exchange (XML)

• 

Test certificate (PDF) - optional

Move to " www.vega.com

" and enter in the search field the serial 

number of your instrument. 

Alternatively, you can access the data via your smartphone:

• 

Download the VEGA Tools app from the " 

Apple App Store

" or the 

Google Play Store

• 

Scan the DataMatrix code on the type label of the instrument or

• 

Enter the serial number manually in the app

Type label

Serial number - Instru-

ment search

Содержание VEGAPULS 66 Profibus PA

Страница 1: ...Quick setup guide Radar sensor for continuous level measurement of liquids VEGAPULS 66 Profibus PA Document ID 47140 ...

Страница 2: ...9 5 Set up with the display and adjustment module 11 5 1 Insert display and adjustment module 11 5 2 Parameter adjustment 12 5 3 Menu overview 15 6 Set up with smartphone tablet PC notebook via Bluetooth 17 6 1 Preparations 17 6 2 Connecting 18 6 3 Sensor parameter adjustment 18 7 Supplement 20 7 1 Technical data 20 Information This quick setup guide enables quick setup and commissioning of your i...

Страница 3: ...mplying with all prevailing regulations and directives The instrument must only be operated in a technically flawless and reliable condition The operator is responsi ble for the trouble free operation of the instrument When measuring aggressive or corrosive media that can cause a dangerous situation if the instrument malfunctions the operator has to implement suitable measures to make sure the ins...

Страница 4: ...ard in field instrumentation The device fulfils the requirements of the following NAMUR recom mendations NE 21 Electromagnetic compatibility of equipment NE 53 Compatibility of field devices and display adjustment components NE 107 Self monitoring and diagnosis of field devices For further information see www namur de 1 7 Radio license for Europe The instrument was tested according to the latest i...

Страница 5: ...y to operate this equipment 1 9 Environmental instructions Protection of the environment is one of our most important duties That is why we have introduced an environment management system with the goal of continuously improving company environmental pro tection The environment management system is certified according to DIN EN ISO 14001 Please help us fulfil this obligation by observing the envir...

Страница 6: ...ss 14 ID numbers instrument documentation 15 Reminder to observe the instrument documentation The type label contains the serial number of the instrument With it you can find the following instrument data on our homepage Product code HTML Delivery date HTML Order specific instrument features HTML Operating instructions and quick setup guide at the time of ship ment PDF Order specific sensor data f...

Страница 7: ...ns 1 Distance from the vessel wall 200 mm the antenna should protrude 10 mm into the vessel 500 mm 19 69 ca 10 mm Fig 2 Distance of the antenna to the vessel wall vessel ceiling 2 Note min socket diameter depending on the socket length For further information see chapter Mounting Mounting ...

Страница 8: ...ut of the cable gland and remove blind plug 4 Remove approx 10 cm 4 in of the cable mantle strip approx 1 cm 0 4 in of insulation from the ends of the individual wires 5 Insert the cable into the sensor through the cable entry 1 2 Fig 3 Connection steps 5 and 6 1 Single chamber housing 2 Double chamber housing 6 Insert the wire ends into the terminals according to the wiring plan Note Solid cores ...

Страница 9: ...cs and connection compartment single chamber housing 1 Voltage supply signal output 2 For display and adjustment module or interface adapter 3 Selection switch for instrument address 4 For external display and adjustment unit 5 Ground terminal for connection of the cable screening 4 3 Wiring plan double chamber housing The following illustrations apply to the non Ex as well as to the Ex ia version...

Страница 10: ...ing to the bus system VEGAPULS 66 Profibus PA 47140 EN 210621 Information Parallel use of an external display and adjustment unit and a display and adjustment module in the connection compartment is not sup ported ...

Страница 11: ...y 90 It is not necessary to interrupt the power supply Proceed as follows 1 Unscrew the housing lid 2 Place the display and adjustment module on the electronics in the desired position and turn it to the right until it snaps in 3 Screw housing lid with inspection window tightly back on Disassembly is carried out in reverse order The display and adjustment module is powered by the sensor an ad diti...

Страница 12: ...y and adjustment module for continuous measured value indication a higher lid with an inspection glass is required 5 2 Parameter adjustment 1 Go to the menu Setup via the display and adjustment module 2 In the menu item Medium you select the medium of your ap plication for example Aqueous solution 3 Select in the menu item Application the vessel the application and the vessel form for example stor...

Страница 13: ...ations is always the seal surface of the thread or flange The function Echo curve memory makes it possible to save the echo curve at the time of setup This is generally recommended and it is absolutely necessary if you want to use the Asset Management functions If possible the curve should be saved with a low level in the vessel With the adjustment software PACTware and a PC a high resolution echo...

Страница 14: ...edium surface because if an incorrect too large value is entered the existing level will be saved as a false signal The level would then no longer be detectable in this area If a false signal suppression has already been saved in the sensor the following menu window appears when selecting False signal suppression Delete An already created false signal suppression will be com pletely deleted This i...

Страница 15: ... 000 m d 100 00 Min adjustment 35 m 0 00 AI FB1 Channnel PV lin perc AI FB1 scaling unit Height AI FB1 scaling 0 00 lin 0 00 100 00 lin 100 00 AI FB1 damping 0 s Lock adjustment Released Menu item Default setting Language Order specific Displayed value Filling height in Backlight Switched on Menu item Parameter Default setting Device status Peak value indi cator Distance Electronics tem perature f...

Страница 16: ... in m Temperature in C Unit SV2 Distance in m False signal suppression Linearisation Linear PIN Date Time Actual date Actual time Reset Copy instrument settings Menu item Parameter Device name VEGAPULS 6 Instrument version Hardware and software version Date of manufacture Date Profibus ident number 4 digit hexadecimal Instrument features Order specific characteristics Additional adjustments Info ...

Страница 17: ...tooth active Off Bluetooth not active The security concept of Bluetooth operation absolutely requires that the default setting of the sensor PIN be changed This prevents unau thorized access to the sensor The default setting of the sensor PIN is 0000 First of all you have to change the sensor PIN in the adjustment menu of the sensor e g to 1111 1 In the adjustment menu go to Additional adjustments...

Страница 18: ... Bluetooth and start the search function The device auto matically searches for Bluetooth capable devices in the vicinity The message Instrument search running is displayed All devices found are listed in the operating window The search is automatically continued continuously Select in the device list the requested device The message Con necting is displayed For the first connection the operating ...

Страница 19: ...19 6 Set up with smartphone tablet PC notebook via Bluetooth VEGAPULS 66 Profibus PA 47140 EN 210621 Fig 10 Example of an app view Setup sensor adjustment App view ...

Страница 20: ...IP68 0 2 bar Options of the cable entry Ʋ Ʋ Cable entry M20 x 1 5 NPT Ʋ Ʋ Cable gland M20 x 1 5 NPT cable ø see below table Ʋ Ʋ Blind plug M20 x 1 5 NPT Ʋ Ʋ Closing cap NPT Material ca ble gland Material seal insert Cable diameter 4 5 8 5 mm 5 9 mm 6 12 mm 7 12 mm 10 14 mm PA NBR Brass nickel plated NBR Stainless steel NBR Wire cross section spring loaded terminals Ʋ Ʋ Massive wire stranded wire 0...

Страница 21: ...21 Notes VEGAPULS 66 Profibus PA 47140 EN 210621 ...

Страница 22: ...22 Notes VEGAPULS 66 Profibus PA 47140 EN 210621 ...

Страница 23: ...23 Notes VEGAPULS 66 Profibus PA 47140 EN 210621 ...

Страница 24: ...ning scope of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior notice VEGA Grieshaber KG Schiltach Germany 2021 Phone 49 7836 50 0 E mail info de vega com www vega com ...

Отзывы: