background image

16

5 Set up with the display and adjustment module

VEGAPULS 64 • 4 … 20 mA/HART - two-wire

51462-EN-160111

100%

0%

0,5 m (19.68

")

5 m

(196.9

")

2

1

3

Fig. 11: Parameter adjustment example min./max. adjustment
1  Min. level = max. measuring distance

2  Max. level = min. measuring distance

3  Reference plane

If these values are not known, an adjustment with the distances of e.g. 

10 % and 90 % is possible. Starting point for these distance specifica-

tions is always the sealing surface of the thread or flange. You can find 

specifications on the reference plane in chapter "

Technical data

". The 

actual level is calculated on the basis of these settings.
The actual product level during this adjustment is not important, 

because the min./max. adjustment is always carried out without 

changing the product level. These settings can be made ahead of 

time without the instrument having to be installed.

The function "

Setup

" allows the echo curve to be saved at the time of 

setup.
Information:

This is generally recommended, however, for use of the Asset 

Management functions it is absolutely necessary. Saving should be 

carried out with a very low level.
The function "

Echo curve memory

" allows up to ten individual echo 

curves to be stored, for example to detect the measurement behav-

iour of the sensor in different operating conditions.
With the adjustment software PACTware and the PC, the stored echo 

curves can be displayed with high resolution and used to recognize 

signal changes over time. In addition, the echo curve saved during 

setup can also be displayed in the echo curve window and compared 

with the current echo curve.

Diagnostics - Echo curve 

memory

Содержание VEGAPULS 64

Страница 1: ...Quick setup guide Radar sensor for continuous level measurement of liquids VEGAPULS 64 4 20 mA HART two wire Document ID 51462...

Страница 2: ...housing 10 5 Set up with the display and adjustment module 5 1 Insert display and adjustment module 12 5 2 Parameter adjustment Quick setup 13 5 3 Parameter adjustment Extended adjustment 15 6 Supple...

Страница 3: ...tions This is a state of the art instrument complying with all prevailing regulations and guidelines The instrument must only be operated in a technically flawless and reliable condition The operator...

Страница 4: ...302372 1 2 V1 2 1 2011 02 for use in closed vessels For operation inside of closed vessels the following conditions must be fulfilled The instrument must be permanently mounted on a closed vessel made...

Страница 5: ...safety VEGAPULS 64 4 20 mA HART two wire 51462 EN 160111 Please help us fulfill this obligation by observing the environmental instructions in this manual Chapter Packaging transport and storage Chap...

Страница 6: ...12 Reminder to observe the instrument documentation The type label contains the serial number of the instrument With it you can find the following instrument data on our homepage Product code HTML Del...

Страница 7: ...clination in steps 2 For angle of inclination infinitely variable Depending on the selected variant the sensor can be rotated in the strap Single chamber housing Angle of inclination in three steps 0...

Страница 8: ...ese can be suppressed by an appropriate adjustment see chapter Setup If you cannot maintain this distance you should carry out a false signal storage during setup This applies particularly if buildup...

Страница 9: ...s To do this lift the terminal block with a small screwdriver and pull it out When reinserting the terminal block you should hear it snap in Proceed as follows 1 Unscrew the housing lid 2 If a display...

Страница 10: ...eased the terminal closes again You can find further information on the max wire cross section under Technical data Electromechanical data 7 Check the hold of the wires in the terminals by lightly pul...

Страница 11: ...mA 2 3 4 1 Fig 8 Electronics and terminal compartment single chamber housing 1 Voltage supply signal output 2 For display and adjustment module or interface adapter 3 For external display and adjustme...

Страница 12: ...ed by 90 It is not necessary to interrupt the power supply Proceed as follows 1 Unscrew the housing lid 2 Place the display and adjustment module on the electronics in the desired position and turn it...

Страница 13: ...her lid with an inspection glass is required 5 2 Parameter adjustment Quick setup To quickly and easily adapt the sensor to the application select the menu item Quick setup in the start graphic on the...

Страница 14: ...nu item you enter the height of the vessel and hence the active measuring range 6 Max adjustment In this menu item you carry out the max adjustment Enter the measuring distance for 100 filling 7 Min a...

Страница 15: ...ghting Diagnosis Information for example on device status peak value simulation echo curve Additional adjustments Date Time reset copy function scaling current output false signal suppression lineariz...

Страница 16: ...djustment is always carried out without changing the product level These settings can be made ahead of time without the instrument having to be installed The function Setup allows the echo curve to be...

Страница 17: ...ent Max adjustment 0 000 m d 100 00 Min adjustment 30 m 0 00 Damping Integration time 0 0 s Current output Current output Mode Output characteristics 4 20 mA Reaction when malfunctions occur 3 6 mA Cu...

Страница 18: ...caling Scaling size Volume in l Scaling format 0 corresponds to 0 l 100 corresponds to 0 l Current out put 1 Current output Meas vari able Lin percent Level Current output Adjustment 0 100 correspond...

Страница 19: ...ed Proceed as follows 1 Select with the menu item False signal suppression and confirm with OK 2 Confirm again with OK 3 Confirm again with OK 4 Confirm again with OK and enter the actual distance fro...

Страница 20: ...hes the metrological conditions in the vessel Extend is used to extend an already created false signal suppres sion This is useful if a false signal suppression was carried out with too high a level a...

Страница 21: ...anded wire with end sleeve 0 2 1 5 mm AWG 24 16 Voltage supply Operating voltage UB Non Ex instrument 12 35 V DC Ex d instrument 12 35 V DC Ex ia instrument 12 30 V DC Ex d ia instrument 17 35 V DC Op...

Страница 22: ...22 Notes VEGAPULS 64 4 20 mA HART two wire 51462 EN 160111...

Страница 23: ...23 Notes VEGAPULS 64 4 20 mA HART two wire 51462 EN 160111...

Страница 24: ...g scope of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without...

Отзывы: