11 Diagnostics and servicing
VEGAMET 841 • 4 … 20 mA
58864-EN-200401
Error code
Cause
Rectification
S016
Adjustment:
Min./Max. ex-
changed
The min./max.
adjustment
points were ex-
changed.
Carry out the adjustment again, correct
the min. and max. values.
S017
Adjustment:
Span too small
Adjustment span
too small
Carry out a fresh adjustment and in-
crease the distance between min./max.
adjustment
S021
Scaling: Span
too small
Scaling span too
small
Carry out a fresh scaling, increase the
distance between min./max. scaling.
S022
Scaling: Scaling
value too large
Value for one
of the two scal-
ing points is too
large
Carry out the scaling again, correct the
min. and max. values.
S062
Pulse priority too
small
Pulse priority too
small
Increase under "
Output
" the entry
"
Pulse output all
" so that max. one pulse
per second is output.
S110
Relay: Span too
small
Relay switching
points too close
together
Increase the difference between the two
relay switching points
S111
Relay: Switching
points inter-
changed
Relay switching
points inter-
changed
Change relay switching points for "
On/
Off
"
S115
Pump control:
Faulty interfer-
ence behaviour
Several relays
are assignef to
the pump control
which are not set
to the same fail-
ure mode
All relays which are assigned to the
pump control must be set to the same
failure mode
S116
Pump control:
Operating mode
faulty
Several relays
that are not con-
figured with the
same mode are
assigned to the
pump control
All relays which are assigned to the
pump control must be set to the same
mode
Error code
Cause
Rectification
C029
Simulation
Simulation active Finish simulation
11.4 Software update
An update of the device software is possible via the Bluetooth inter-
face.
The following components are required:
Out of specification
Function check