26
9 Supplement
VEGACAP 69 • Two-wire
31177-EN-230510
9 Supplement
9.1 Technical data
Note for approved instruments
The technical data in the respective safety instructions which are included in delivery are valid for
approved instruments (e.g. with Ex approval). These data can differ from the data listed herein, for
example regarding the process conditions or the voltage supply.
All approval documents can be downloaded from our homepage.
General data
Material 316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
Ʋ Process fitting - flange
PP or PTFE
Ʋ
Insulation (fully insulated)
FEP
Materials, non-wetted parts
Ʋ
Probe - double rod fully insulated:
ø 14 mm (0.551 in)
316L
Ʋ
Plastic housing
Plastic PBT (Polyester)
Ʋ
Seal between housing and housing lid Silicone
Ʋ
Ground terminal
316L
Ʋ
Cable gland
PA, stainless steel, brass
Ʋ
Sealing, cable gland
NBR
Ʋ
Blind plug, cable gland
PA
Process fittings
Ʋ
Flanges
DIN from DN 50, ASME from 2"
Weight
1)
Ʋ
Instrument weight
0.8 … 4 kg (0.18 … 8.82 lbs)
Ʋ
Rod weight: ø 14 mm (0.551 in)
2000 g/m (22 oz/ft)
Sensor length (L)
0.2 … 4 m (0.656 … 13.12 ft)
Max. lateral load
10 Nm (7.4 lbf ft)
Torque for NPT cable glands and Conduit
tubes
max. 10 Nm (7.376 lbf ft)
Measuring frequency
430 kHz
Output variable
Output
Two-wire output
Suitable controllers
VEGATOR 141, 142, 620, 621, 622
Output signal
> 4 … < 20 mA (not standardised)
Fault message
< 2.3 mA
Switching delay
Ʋ
When immersed
0.7 s
1)
Flange weight not considered