17
3 Mounting
SHLD1 • Source holder
62092-EN-190704
Mounting brackets, angled attachments as well as mounting clamps
are available for mounting.
Vertical pipeline, 30° inclined, diameter 50 … 100 mm
(1.97 … 3.94 in)
For small pipe diameters 50 … 100 mm (1.97 … 3.94 in), a diagonal
radiation path is recommended. The distance of the beam through the
medium is thus longer and an improved measuring effect is achieved.
In such cases, the optional lead shielding for the detector is recom-
mended in order to avoid influence from secondary radiation sources.
Fig. 9: 30° measurement setup on a pipeline with diameter 50 … 100 mm
(1.97 … 3.94 in)
Vertical pipeline, diameter 50 … 600 mm (1.97 … 23.62 in)
For pipe diameters 50 … 600 mm (1.97 … 23.62 in), a straight radia-
tion path is possible. The radiometric sensor can be mounted either
horizontally or vertically.