Important Information

Please Read Before Use 

 

12

 / 

20

                                                                                                File No.: IR-IM-001

 

The device measure result is not accurate. 

 

The measure sensor is dirty, and need to 
use a clean soft cloth or cotton bud that 
can be moistened with 70% alcohol. 

 

The patient uses the device 
immediately after exercising or walking a 
long time. And need to rest for 5 mins to 
10mins to measure the temperature. 

The device displays the symbol”Err”. 

The operation environment temperature 
changes suddenly, and the temperature 
difference is greater than 8

, the device will 

display ”Err”. And it just needs to put the 
device in the operation environment for 
20mins and more, after that to use the device. 

10.

 

Safety precaution:   

——

Indicates  a  potentially  hazardous  situation  which,    if  not  avoided,  could  result  in 

serious injury.

 

Read all instructions carefully before use. 

 

Warning: 

 

 

No modification of this equipment is allowed. 

 

Check the thermometer for signs of damage or wear before each use. Damaged 

or worn thermometer should not be used. 

 

Under  any  circumstances,  the  temperature  taking  result  is  ONLY  for  reference. 

Before taking any medical action, please consult your physician. 

 

MR UNSAFE: The device cannot be used in an MR environment. 

 

The user should report any serious incident that has occurred in relation to the 

device  to  the  manufacturer  and  the  competent  authority of the  Member  State  in 
which the user and/or patient is established. 

10-1. Product cautions: 

 

If the sensor is degraded or loosened, and the measure value is incorrectly, please stop using 

the device immediately and contact your equipment provider.   

 

To avoid electrical shock: Keep the unit away from water. Do not immerse the unit in liquid. 

Do not use while bathing. Do not reach for a unit that has fallen into water. 

 

Never operate the unit if it  has any damaged parts, if it has been dropped or submersed in 

water. Promptly send it to a service center for examination and repair.   

 

If any abnormality occurs, discontinue use until the unit has been examined and repaired.   

Содержание IR-08MT

Страница 1: ...Read Before Use 1 20 File No IR IM 001 VEGA TECHNOLOGIES INC INSTRUCTION MANUAL Non contacted Infrared forehead thermometer Model No IR 08MT Revision V1 0 Date of issue 2022 7 22 Read this instruction...

Страница 2: ...evice 3 3 Product description 4 4 Product features 5 5 Device specifications 5 6 Operating Instructions 7 7 Cleaning Disinfecting 10 8 Storage Maintenance Repair 10 9 TROUBLESHOOTING 11 10 Safety prec...

Страница 3: ...mperature gun for measuring human body temperature The product is widely used in schools customs hospitals families Easy to use mode selection The Non contacted Infrared forehead thermometer is intend...

Страница 4: ...Please Read Before Use 4 20 File No IR IM 001 3 Product description 3 1 Visual appearance and structure LED ON OFF button Measure temperature button Sensor Enclosure Battery slot Positive pole Negati...

Страница 5: ...oltage indication With AAA size LR03 alkaline battery 2pcs Optional storage case and carrying bag for easy storage 5 Device specifications NO Item Specification 1 Operation mode Continuous mode 2 Meas...

Страница 6: ...erature mode One Bi emits when body temperature 37 5 Six Bi emits when body temperature 37 5 16 Automatic shut off Stop operating the device it wills automatic shut off after 20 5 seconds 17 Low power...

Страница 7: ...zer At this time the measurement can start Press the button to start the measurement B Switch off After startup press M button and then the unit turn off in the Direct Mode After startup the unit will...

Страница 8: ...M button and the device switch off Then operated once again according the chapter adjust mode 6 1 4 Memories recall When the device is startup and displays on the screen the device defaults to display...

Страница 9: ...the result will display in the LED screen If you need to measure the person temperature please choice the adjust mode and the LED screen will display the symbol Please see the chapter 6 1 3 Direct mod...

Страница 10: ...lens and the LED Do not put the device into the water cleaning in order to avoid damage to the device use cotton swabs dipped in alcohol or using alcohol content of 70 cotton cloth or cotton to cleani...

Страница 11: ...he patient or the user can replace of batteries and the children use or replace the battery need under the adult supervision 8 3 Repair Never open or attempt to repair the device by yourself as otherw...

Страница 12: ...ter for signs of damage or wear before each use Damaged or worn thermometer should not be used Under any circumstances the temperature taking result is ONLY for reference Before taking any medical act...

Страница 13: ...es that can be swallowed Keep children away from packaging materials risk of suffocation Store the parts in a clean location for avoiding infection Do not use the device after exercising and need to r...

Страница 14: ...o as not to damage the body of the unit 10 4 Cleaning Cautions Use the cotton buds to clean the probe inside when it has dirt Do not use water or fingers to contact with the probe directly Do not imme...

Страница 15: ...ted to misuse improper operation accidental damage or unauthorized repairs and is not extended to charges of or for labor repairs All items returned must be properly packaged and shipped prepaid by th...

Страница 16: ...w The customer or the user of the device should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The device us...

Страница 17: ...ins power quality should be that of a typical commercial or hospital environment If the user of the device requires continued operation during power mains interruptions it is recommended that the devi...

Страница 18: ...to 800MHz d 2 3 P 800MHz to 2 5GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in m...

Страница 19: ...er frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Table 5 refer...

Страница 20: ...Important Information Please Read Before Use 20 20 File No IR IM 001 MedPath GmbH Mies van der Rohe Strasse 8 80807 Munich Germany 0123...

Отзывы: